Page 14
ecom-B
5. Mean value (option)
By mean value measurement function, measurements can be sam-
pled within an adjustable time frame and mean values can be calcu-
lated. Should the several measurement values or the mean value
result be stored a storage place has to be selected as described in
chapter 4.
Once the fresh air calibration is com-
pleted, select the menu point
„Mean
values“
. Before starting, the parame-
ters
„Measurement time“
and
„Scan-
ning“
should be checked or modified if
need be. The meaning is respectively:
- Measurement time
= Time frame
during which the mean values will
be sampled
- Scanning
= Time interval between the
measurements considered for mean
value calculation
„Measurement time“
and
„Scanning“
can be adjusted as follows:
- select menu point and confirm with
<Enter>
- set the desired time using the cursor keys
After adjustment / control the measurement can be started by
<Start> / <Enter>
. With the cursor keys
<up/down>
, it is possible to
switch between current and average readings. The measurement
can be interrupted by pressing
<Standby>
(pressing
<Standby>
again sets the measurement continued).
Comb.measure.
Mean value
Atmosph. boiler
Burn. w. blower
Diagnosis gas
Pressure tests
Data processing
Adjustment
Control
Mean value
Start
Meas. time
Scanning
Continue with :
↑↓
↵
↵↵
↵