User guide 

AUDI

O

SMART 

ECH001XN141-A4 – 07/2022

 

 

33 

Chapter 5. The use of the software ECHOSOFT 

5.4 Settings 

 

 

5.4.2

 

Medical software 

This  section  is  used  to  configure  a  third-party  patient 
management software to import audiometry curves. 
 
A  first  drop-down  menu  lets  you  select  the  software 
used.  Next,  you  must  define  the  location  where  the 

ECH

O

SOFT

 

software  will  retrieve  the  patient  infor-

mation. Finally, you need to define the location where 
the 

ECH

O

SOFT

 

should store the results once the meas-

urement  is  completed,  so  that  the  third-party  software 
can retrieve the curves. 
 

5.4.3

 

Puretone audiometry settings 

In this section you can select the frequency available for the pure tone audiometry and define the setting for the automatic 
mode. 

 

1.

 

Select the frequencies available in the audiometric measurement 

 

2.

 

The automatic threshold measurement mode searches for a patient's hearing threshold over the range of frequencies 
preselected  in 

1

.  The  frequencies  are  scanned  from  1000Hz  to  the  highest  frequency,  then  from  1000Hz  to  the 

lowest  frequency.  The  repetition  of  the  test  at  1000Hz  depends  on  the  selection  of  the  "

double  validation  at 

1000Hz

" box. For each frequency, the test starts at the selected "

starting power

". The automatic algorithm per-

forms the power changes according to the rising threshold method, respecting the "

maximum power

" and "

mini-

mum power

" settings. The "

Response  Time

" is the time limit after the stimulus presentation during which the 

patient's response is considered valid. The intervals between two stimuli are randomly modified according to the 
maximum and minimum intervals set. 

3.

 

The modifications must be validated by clicking on 

« Save »

 

 

The maximum frequency during testing may be limited depending on the stimulator (headphones) used. 
For stimulation above 8000Hz, you need the "HF Audiometry" module and a high frequency headset. 

Summary of Contents for AUDIOSMART

Page 1: ...a brand of lectronique du Mazet ZA Route de Tence 43520 Le Mazet Saint Voy FRANCE T l 33 4 71 65 02 16 Email contact electroniquedumazet com Firmware 1 8 0x Web www electroniquedumazet com Software 2...

Page 2: ...ntil it is destroyed This equipment has been designed and manufactured for use in otologic diagnosis Use is restricted to professionals who have undergone appropriate training If you have a problem or...

Page 3: ...MART 11 2 1 Initial Start up of the device 11 2 1 1 Switching on starting 11 2 1 2 Touch screen calibration 11 2 1 3 Password 11 2 1 4 Home screen 12 2 1 5 Switching off the device 12 2 2 General devi...

Page 4: ...olor code 40 5 8 2 Automatic mode audiometry with Auto mode masking 40 5 8 3 Calculation method 40 5 9 Merge measurement 41 5 10 Using keyboard shortcuts 42 5 11 Use of the microphone 42 6 Maintenance...

Page 5: ...third parties including patients 1 2 Presentation of the device AUDIOSMART is designed for screening documentation monitoring and diagnosis of hearing functions It is intended for ENT doctors audiolo...

Page 6: ...seek advice in case of doubt 1 2 5 Side effects No side effects identified to date 1 2 6 Units of measurement For all these devices the units of measurement are expressed in the units of the internati...

Page 7: ...e must be connected to a computer with a certified medical power supply double insulation according to ISO 60601 1 CAUTION No modifications to the device are permitted It is strictly forbidden to open...

Page 8: ...ial use case No specific cases identified See section Erreur Source du renvoi introuvable for contraindications 1 5 Commissioning Check that the device is not damaged if you have any doubts about the...

Page 9: ...Acoustic stimulator connection for audiometry and electrophysiology OAE probe connection for otoacoustic emissions Headset connection Back side Warning this logo draws your attention to a specific poi...

Page 10: ...label Direct current Serial number Manufacturer Year of manufacture Country of production Product reference CE marking 1 7 Identification label Information and specifications are given on the back of...

Page 11: ...so it is the responsibility of the practitioner to ensure that the following recommendations are followed 1 9 1 Good practices for computer security Keep your software up to date including the operati...

Page 12: ...1 3 Password After the screen calibration the password definition windows appear If you choose to set up a password it will be asked to you each time you start the device To do this click on Lock the...

Page 13: ...em language by clicking on the flags at the bottom left of the screen At the bottom right the serial number of your device is displayed A title bar is present on all windows of the device From left to...

Page 14: ...button informs you about the hardware and software version of the device and the amount of free memory space on the AUDIOSMART device The Data factory reset button resets the measure ment parameters...

Page 15: ...nit click on Config The settings window will open Click on Headset to access the selection of the headset that will be recognized when connected to the jack Select the headset model you will use and c...

Page 16: ...nterpretation of the resulting audiogram allows to know the degree of hearing loss and the kind of deafness Pure tone audiometry also allows to find a discomfort threshold or also the research for any...

Page 17: ...hen connect the ear phone Mini DIN to the Audio connector of the AUDIOSMART unit For bone conduction measurement put the vibrator on the mastoid or on the forehead for Webber test and put a foam ear t...

Page 18: ...st click on the desired field to have the keyboard appearing on the screen It is possible to use a digital keypad by clicking on the 123 key at the bottom left Here the information about the patient i...

Page 19: ...fying glass icon at the bottom of the screen enables to search for a patient thanks to his her name or surname To select a patient click on the corresponding line This opens a new page summarizing inf...

Page 20: ...ce you select the Pure Tone Audiometry test you can choose among four modes of diagnosis Automatic patient mode Automatic doctor mode Manual doctor mode Weber Mode 4 2 1 1 Patient mode The patient mod...

Page 21: ...User guide AUDIOSMART ECH001XN141 A4 07 2022 20 Chapter 4 Handheld mode measurement 4 2 Audiometry...

Page 22: ...ends the curve is constructed You have the choice between saving the curve with the Save button or delete it by leaving the window with the back cross 4 2 1 2 Automatic doctor mode Automatic doctor m...

Page 23: ...stimuli Once the acquisition protocol ends the curve is constructed You have the choice between saving the curve with the Save button or delete it by leaving the window with the back cross 4 2 1 3 Man...

Page 24: ...h stimulation indicate if the patient hears the sound at the left right or on both sides You can see the curve at any time by pressing the Graph button Then you can save the curve by clicking on Save...

Page 25: ...of the test the device displays the words from the list on the screen This list is randomly chosen to guarantee a reliable test without the risk of the patient learning the words The Settings button a...

Page 26: ...e test is finished click on Graph to see the curve You can then save it or make other Pure Tone audiom etry or Speech audiometry measurements This window lets you select the eat and the power Click St...

Page 27: ...audiometric normal Speech audiometry The abscissa scale represents the power of the stimuli in dB HL The ordinate scale represents the percentage of correctly repeated words The last image on the righ...

Page 28: ...lass II compliant with EN 60601 1 5 2 Installation 5 2 1 Installing the software The ECHOSOFT is supplied as an executable software file designed for automatic installation of the application on your...

Page 29: ...all a USB driver to have this functionality Since the version 2 1 6 0 the USB driver installation is automati cally trigger during the ECHOSOFT installation process but for some reasons you could have...

Page 30: ...All information related to a patient can be modified To access the patient record screen select the patient and click on the Edit Patient s Profil at the bottom of the main screen 5 3 2 Import patient...

Page 31: ...choosing the corresponding patient or simply creating a new one 5 3 2 2 Synchronize all patients with the database This option lets you add all of the AUDIOSMART patients to the ECHOSOFT database The...

Page 32: ...atient se lected in the list The Delete button is used to defini tively delete the patient from the ECHOSOFT database 5 3 3 2 Delete a patient from AUDIOSMART A patient can be deleted from the AUDIOSM...

Page 33: ...abase located in another folder It is possible to select a database located on your computer on a USB key or a network shared folder Relocating the database move the current database to another folder...

Page 34: ...reshold measurement mode searches for a patient s hearing threshold over the range of frequencies preselected in 1 The frequencies are scanned from 1000Hz to the highest frequency then from 1000Hz to...

Page 35: ...ributes to the improvement of your products To take advantage of these updates check our website http echodia fr en downloads regularly to see if the latest version available corresponds to your curre...

Page 36: ...omputer PC or Mac Start the ECHOSOFT the window below appears Connect the device to the computer and click on the USB button from the home screen of the AUDIOSMART The Audiometer button 3 becomes avai...

Page 37: ...try or Weber test in the case of a bone conduction 4 Start automatic mode See subsection 5 10 for the settings When the Patient Mode box is checked the operator no longer controls the test it is only...

Page 38: ...ation is in progress Green no stimulation Red stimulation is running 11 Save or create a measurement 12 The cross represents the current position of the mouse cursor make a left click to start the sti...

Page 39: ...automatic mode 9 The Headphones button allows you to see which stimulator is active and to switch between the two audio outputs Thus it is possible to connect the headphones and the bone vibrator eac...

Page 40: ...User guide AUDIOSMART ECH001XN141 A4 07 2022 39 Chapter 5 The use of the software ECHOSOFT 5 7 Exploitation sur ECHOSOFT...

Page 41: ...wer right ear The blue cross with a bottom right arrow the sound was played but the patient didn t answer left ear 2 Summary table of standard audiometric indicators 3 Weber test display area Abscissa...

Page 42: ...ight or left the stimulus air bone or speech and the frequency The masking in Auto Mode is calculated by applying the differential to the stimulation power sent Thus it varies with each change in stim...

Page 43: ...eadset 40dB Therefore the need for masking and its value may vary depending on the stimulator used which is automatically identified by the module In order to calculate the AC masking the AC and BC th...

Page 44: ...raph the curves will be shown with their respective dates and times In the example above only the right side has two measurements By showing two measurements at the same time in addition to being able...

Page 45: ...one setup is done thanks to the third tab at the top left of the audiometry window It is possible to select the input device the device list will depend on the computer and the sound card It is possib...

Page 46: ...e of the software ECHOSOFT 5 11 Use of the microphone To use the microphone keep the A key pressed and speak the light indicator at the bottom left then turns red The volume is indicative it may vary...

Page 47: ...ty it is imperative to follow the operating instructions provided in this manual AUDIOSMART devices are designed to have a service life of 5 years 6 2 Cleaning 6 2 1 Device case The device case only r...

Page 48: ...nected Connecting the stimulator Stimulator is out of order Contact your distributor to initiate the after sales service procedure Gas and or liquid leakage from the case during operation or not Batte...

Page 49: ...other The warranty is void if the operating instructions in this manual are not strictly followed The warranty period is 24 months from the date of delivery of the device Transport and packaging cost...

Page 50: ...ture 6 5 Disposal As soon as any deterioration is detected the product must be cleaned with a broad spectrum disinfectant and returned to the manufacturer If the product becomes inoperative or unusabl...

Page 51: ...from a computer or the power adapter see Erreur Source du renvoi introuvable Resolution 320 x 240 65000 colours Touch screen Resistive screen usable with a finger or with a stylus Energy comfort Back...

Page 52: ...lection Central frequency Hz Narrow band masking noise AC BC Lower cut off Hz Upper cut off Hz Maximum power dB EM min 10 dB EM Maximum power dB HL min 10 dB HL Maximum power dB HL min 10 dB HL 125 11...

Page 53: ...t least 30 Rapid transients in bursts IEC 61000 4 4 2 kV for lines power supply electric 1 kV for lines input output 2 kV for power lines The quality of the power supply should be that of a typical co...

Page 54: ...nnot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an on site electromagnetic investi gation should be considered If the field strengt...

Page 55: ...otherapy depressotherapy and elec trotherapy urological rehabilitation equipment Electronique du Mazet also owns the Echodia brand which has a dedi cated R D office specialized in functional explorati...

Page 56: ...User guide AUDIOSMART ECH001XN141 A4 07 2022 Your dealer distributor ELECTRONIQUEDU MAZET ZA ROUTE DE TENCE 43520 LE MAZET SAINT VOY T l 33 4 71 65 02 16 Mail sav electroniquedumazet com...

Page 57: ...care etc I have read the instruction manual and noted the warranty and service conditions In the event that Electronique du Mazet or its distributors do not receive this form duly completed and signed...

Reviews: