ADJUSTMENTS AND REPLACEMENTS
46
01JUN99 – 5B6331
Removing the HEAT EXCHANGER
Warning
Dangerous Voltage
[1] De-energize the PROCESSOR.
[2] Lift the TOP COVER.
[3] Remove:
• DRIVE END PANEL
• WET SECTION COVER
• EXIT RACK
• DRYER RACK
[4] Install 2 CLAMPS TL-2170 on the HEAT EXCHANGER FORMED TUBING.
[5] Remove the 2 HOSE CLAMPS on the HEAT EXCHANGER FORMED TUBING.
[6] Remove:
• FORMED TUBING from the ends of the BARBED FITTINGS
• BARBED FITTINGS from the HEAT EXCHANGER
• O-RING SEALS
• HEAT EXCHANGER from the WASH TANK
H150_0150HA
H150_0150HCA
HOSE CLAMP
BARBED FITTING
O-RING SEAL
EXCHANGER
HEAT