ADJUSTMENTS AND REPLACEMENTS
Additional Service Procedures
06DEC05
AR4366-1
Page
206 of 219
Important
You must use a LEAD SHEET or a LEAD APRON under the DOSIMETER. If a LEAD SHEET
or a LEAD APRON is not used, the values on the DOSIMETER will be false by a maximum
of 20%.
3
Place a LEAD SHEET or a LEAD APRON and the DOSIMETER TL-4818 in the center of
the area.
4
Move:
•
POWER SWITCH on the DOSIMETER to the “ON” position.
•
READ/RESET SWITCH to the “RESET” position and then to the “READ” position.
5
Check that “0.00” displays on the LCD DISPLAY.
6
Ask the RT to expose the DOSIMETER.
7
Check that the DOSIMETER displays a value of “20.0 mR”
±
“2.0 mR” on the LCD
DISPLAY.
H149_0015BC
ON
OFF
READ
RESET
LCD DISPLAY
POWER
SWITCH
H149_0015BCA
READ/RESET
SWITCH