Resolving power
1mA
Test time
1s~16h
measuring range
Voltage
readback value
range
0~19.999V
0 ~150.00V
0~19.999V
0~500.00V
Resolving power
1mV
10mV
Accuracy
±(0.05%+0.1%FS)
Current
readback value
range
0~3.000A
0
~
40.00A
0~3.000A
0~20.00A
0~3.000A
0~15.00A
Resolving power
1mA
10mA
Accuracy
±(0.05%+0.1%FS)(0~30A)
±(0.05%+0.1%FS)
±(0.08%+0.1%FS)(30~40A)
Power
readback value
range
400W
200W
400W
Resolving power
10mW
Accuracy
±(0.1%+0.5%FS)
Scope of protection
Overvoltage protection
Overvoltage protection greater than 155V
Overvoltage
protection greater
than 520V
Overcurrent protection
More than 45A cut off input
More than 22A cut
off input
More than 16A cut
off input
Over power protection
420W
220W
420W
Overtemperature protection
85
℃