When you switch on

 

 and the home page opens, 3 menus will be available:

1

 “Treatment of meibomian blepharitis” menu. 

See section on “TREATMENT OF MEIBOMIAN BLEPHARITIS USING

 

”.

2

 “Consumables”  menu. 

This menu shows the number of flashes remaining on the activation card and its 

optic

 

cartridge. once the flash credit is exhausted, user only needs to replace the 

 

activation card and its associated 

optic

 cartridge.

3

 “Configuration” menu.

 

This menu will allow you to access the following 4 sub-menus:

PIN CODE

 (screen K)

This code may be changed to a customized number.The default PIN code is “0000”.  If you 

do not change this code, you will not be requested to enter the PIN upon the start-up of the 

device.  However, if you change the default PIN code, you will be required to enter your PIN 

whenever

 

 is restarted.

To change the PIN code:

- Go to the “Configuration” menu.

- Select the “PIN code” sub-menu.

Using the digital keypad:

- Enter the current PIN code, and validate using the “V” key.

- Enter the new PIN code and validate using the “V” key.

- A message indicating that the codes are valid will appear.  Click on “OK”.

- The “Configuration” menu will be displayed.

It is possible to return to the “Configuration” menu at any time by clicking on icon 

7

.

The “C” key can be used to correct the figure in each box.

SELECT LANGUAGE

 (screen 

L

)

Click on the flag which corresponds to the required language.

The “Language change confirmed” message will be displayed; click on “OK” to confirm 

the change. The “Choice of language” sub-menu will be displayed once more.

The return icon 

7

 can be used to return to the “Configuration” menu.

SETTING OF PARAMETERS (VOLUME AND BRIGHTNESS)

 

(screen 

M

)

It is possible to adjust:

- The volume, using the 2 arrows associated with the loudspeaker icon.

-  The brightness of the touch screen, using the 2 arrows associated with the light bulb 

icon.

The return icon 

7

 can be used to return to the “Configuration” menu.

MAINTENANCE

This access is exclusively reserved for the after-sales service.

IMPORTANT

:

 We would strongly advise you to make a note of your PIN code.  If you have entered a PIN code which you cannot 

remember, it will be necessary to return your 

 

 to us before it can be used again.  Only our technical service is qualified to 

execute the requisite operations for the unlocking of your

 

. This service is available with the device out of guarantee, and 

will be invoiced accordingly.

M

7

445

15

16

17

18

L

7

445

K

7

NOTE:

 

• After 7 seconds of inactivity, 

 screen will switch over to standby mode. Flashing is still possible. Touch the screen to reactivate.

• After around fifteen minutes of inactivity, 

 will switch over to pause mode. Press the 

"play" icon 

13

 

to reactivate.

Summary of Contents for E-Eye

Page 1: ...User Manual...

Page 2: ...nging and is a practical example of this fact Your patients can now enjoy the benefits of a new treatment for meibomian blepharitis dry eye syn drome using state of the art pulsed light technology wit...

Page 3: ...ETC The manufacturer s guarantee for parts and labour will only be valid if your device is returned in its complete original packaging All items of original packaging in which your was delivered shou...

Page 4: ......

Page 5: ...onal user of must ensure that only persons who fulfil the criteria for use described in the present manual will receive treatment using Before using your the following information should be read caref...

Page 6: ...lement for less than two months On an injured burned or infected skin On skin lesions of unusual appearance which will require investigation On high risk areas blemishes birthmarks warts unhealed woun...

Page 7: ...nection Activation card slot Applicator holding Ventilation grid Visor to protect touchscreen Mobile head Release clips for the mobile head Mobile head Applicator zone Applicator zone Air vents Trigge...

Page 8: ...use of the device 13 Pause icon for the flash treatment session pressing this icon will place in pause mode In this mode no flash can be emitted from the applicator To exit this mode press the same i...

Page 9: ......

Page 10: ...plicator photo 3 Push the extractor into the mobile head to its full length photo 4 Apply pressure simultaneously to the 2 retaining lugs of the mobile head and remove the mobile head photo 5 Remove t...

Page 11: ...nce entered these codes will not be requested a second time Connect using the mains lead The internal test procedure of will commence and will last for a few seconds Upon the completion of this phase...

Page 12: ...n a generous layer minimum 1cm thick on the cheekbone and the tempo ral areas making sure that the gel does not get under the eye mask Locate your mask goggles and put them on If there are other perso...

Page 13: ......

Page 14: ...tivation card and its optic cartridge must be replaced when the number of flashes reaches zero STAGE 2 COOLING OF OPTICAL PARTS During the procedure for the replacement of the optic cartridge before s...

Page 15: ...tridges are not symmetrical so they must be inserted in the right sense and certainly not in the other Check that the cartridge is correctly engaged PRECAUTIONS throughout the procedure for the replac...

Page 16: ...the old activation card and insert the new card supplied with the optic cartridge in the slot provided for this purpose photo I Each activation card carries the same serial number as the optic cartrid...

Page 17: ...on icon 7 The C key can be used to correct the figure in each box SELECT LANGUAGE screen L Click on the flag which corresponds to the required language The Language change confirmed message will be d...

Page 18: ......

Page 19: ...o danger or entail the risk of irreparable damage to your device Once o p t i c cartridge is removed it should be handled by the plastic parts only above all the electrodes the tube or the reflector s...

Page 20: ...sothiazolinone 0 05 0 15 Deionized water Q S 100 For external use only Not to be used on broken burned or infected skin Keep out of reach of children The tool extractor is used for the separation of t...

Page 21: ...of your device Never apply the clean spray directly to the internal components of the applicator APPLICATOR BODY All plastic components of the applicator must be cleaned using a clean cloth which has...

Page 22: ...ols of the optical parts of the applicator to end SECU 03 SECU 04 SECU 05 SECU 07 SECU 09 Your device is blocked Contact our technical department by filling out the contact form on www esw vision com...

Page 23: ...pheric pressure 70 106 kPa MARQUAGE Reference to the treatment of meibomian blepharitis dry eye syndrome using device Unique serial number of your E SWIN Rue des C tes d Orval ZA de la Pr v t 78550 Ho...

Page 24: ...considered If the measured field strength in the location in which the device is used exceed the applicable RF compliance level above the device should be observed to verify normal operation If abnor...

Page 25: ...or the applicator as you may risk exposure to high voltages Never use cartridges and acces sories other than those designated and sold by E Swin SAS for use with ENVIRONMENTAL PROTECTION You must abi...

Page 26: ...ns described in this manual will invalidate the guarantee The company E Swin SAS will not accept liability for any accident arising from failure to observe the instructions contained in this manual or...

Page 27: ...0197 ZA de la Pr v t Rue des C tes d Orval 78550 Houdan FRANCE 01 30 46 37 61 www esw vision com M029GB000F08A rev F 2021 06 11...

Reviews: