E-Swin E-Eye User Manual Download Page 14

26

27

REPLACING

 

 

      

Optic 

CARTRIDGE

 REPLACING THE OPTIC CARTRIDGE

•  For any initial use, 

 

optic

 cartridge be pre-installed inside the applicator.

•   Each 

optic

 cartridge is associated with an activation card.

•   A 

 device may be used by several persons, each of whom will have their own 

optic

 cartridge. associated with its own activation 

card.

•  Replacement of the 

optic

 cassette must be undertaken with the utmost care. Improper handling may result in permanent damage to 

the applicator, and to 

 in general. The guarantee does not cover improper handling. We therefore recommend that you pay very 

careful attention to the procedure to be observed.

•  You may have occasion to replace the 

optic

 cartridge before the originally scheduled number of flashes has been exhausted, e.g. in case 

of a change of user. where a tube has been broken, where the 

optic

 cartridge is no longer working, etc.

 STAGE 1

CHECKING THE NUMBER OF FLASHES AVAILABLE ON THE ACTIVATION CARD.

You can monitor your consumption of flashes in real time (see icon 

7

 for the indication of the 

number of flashes remaining), or through the «Consumables» menu.
When the flash credit balance reaches zero, the main menu «Treatment of meibomian ble

-

pharitis» will immediately become inactive, and will be greyed-out (screen 

J

).

 

The activation card and its

 

 

optic

 cartridge must be replaced when  

  the number of flashes reaches zero.

 STAGE 2

COOLING OF OPTICAL PARTS.

•  During the procedure for the replacement of the

 

 

optic

 cartridge, before swit-

ching off and disconnecting your device, you must wait for the fans in the applicator to shut 

down (this may take several minutes). This procedure will allow the cooling of the inter

-

nal optical components of the applicator.  

•  Unlike the fans in the applicator, which will shut down after a certain time, the 

fans in the base unit will continue to run until the mains lead of the device is finally 

 

disconnected.

 STAGE 3

TURNING OFF THE DEVICE.

• Switch off the device. 
• Disconnect the mains lead of the device. 

 STAGE 4

SEPARATION OF THE MOBILE HEAD OF THE 

APPLICATOR USING THE 

extractor

 AND 

EXTRACTION OF 

 

optic

 CARTRIDGE IN 

PLACE. 

•  Hold the applicator with

 

 logo facing you 

and the cable which connects the applicator to 

the base unit on your left. Insert the 

extractor

 

between the mobile head and the fixed lateral 

shell of the applicator (photo 

16

).

•   Push  the 

extractor

 into the mobile head to its 

full length (photo 

17

).

•  Apply pressure simultaneously to the 2 retaining 

lugs of the mobile head and remove the mobile 

head (photo

 

18

).

•   Remove  the 

extractor

 and place the mobile 

head carefully on a flat surface (photo 

19

).

•  Identify the guide notches of the cartridge 

(photo 

20

).

•  Carefully extract the cartridge and remove it 

completely (photo 

21

).

16

17

19

21

J

7

18

20

Summary of Contents for E-Eye

Page 1: ...User Manual...

Page 2: ...nging and is a practical example of this fact Your patients can now enjoy the benefits of a new treatment for meibomian blepharitis dry eye syn drome using state of the art pulsed light technology wit...

Page 3: ...ETC The manufacturer s guarantee for parts and labour will only be valid if your device is returned in its complete original packaging All items of original packaging in which your was delivered shou...

Page 4: ......

Page 5: ...onal user of must ensure that only persons who fulfil the criteria for use described in the present manual will receive treatment using Before using your the following information should be read caref...

Page 6: ...lement for less than two months On an injured burned or infected skin On skin lesions of unusual appearance which will require investigation On high risk areas blemishes birthmarks warts unhealed woun...

Page 7: ...nection Activation card slot Applicator holding Ventilation grid Visor to protect touchscreen Mobile head Release clips for the mobile head Mobile head Applicator zone Applicator zone Air vents Trigge...

Page 8: ...use of the device 13 Pause icon for the flash treatment session pressing this icon will place in pause mode In this mode no flash can be emitted from the applicator To exit this mode press the same i...

Page 9: ......

Page 10: ...plicator photo 3 Push the extractor into the mobile head to its full length photo 4 Apply pressure simultaneously to the 2 retaining lugs of the mobile head and remove the mobile head photo 5 Remove t...

Page 11: ...nce entered these codes will not be requested a second time Connect using the mains lead The internal test procedure of will commence and will last for a few seconds Upon the completion of this phase...

Page 12: ...n a generous layer minimum 1cm thick on the cheekbone and the tempo ral areas making sure that the gel does not get under the eye mask Locate your mask goggles and put them on If there are other perso...

Page 13: ......

Page 14: ...tivation card and its optic cartridge must be replaced when the number of flashes reaches zero STAGE 2 COOLING OF OPTICAL PARTS During the procedure for the replacement of the optic cartridge before s...

Page 15: ...tridges are not symmetrical so they must be inserted in the right sense and certainly not in the other Check that the cartridge is correctly engaged PRECAUTIONS throughout the procedure for the replac...

Page 16: ...the old activation card and insert the new card supplied with the optic cartridge in the slot provided for this purpose photo I Each activation card carries the same serial number as the optic cartrid...

Page 17: ...on icon 7 The C key can be used to correct the figure in each box SELECT LANGUAGE screen L Click on the flag which corresponds to the required language The Language change confirmed message will be d...

Page 18: ......

Page 19: ...o danger or entail the risk of irreparable damage to your device Once o p t i c cartridge is removed it should be handled by the plastic parts only above all the electrodes the tube or the reflector s...

Page 20: ...sothiazolinone 0 05 0 15 Deionized water Q S 100 For external use only Not to be used on broken burned or infected skin Keep out of reach of children The tool extractor is used for the separation of t...

Page 21: ...of your device Never apply the clean spray directly to the internal components of the applicator APPLICATOR BODY All plastic components of the applicator must be cleaned using a clean cloth which has...

Page 22: ...ols of the optical parts of the applicator to end SECU 03 SECU 04 SECU 05 SECU 07 SECU 09 Your device is blocked Contact our technical department by filling out the contact form on www esw vision com...

Page 23: ...pheric pressure 70 106 kPa MARQUAGE Reference to the treatment of meibomian blepharitis dry eye syndrome using device Unique serial number of your E SWIN Rue des C tes d Orval ZA de la Pr v t 78550 Ho...

Page 24: ...considered If the measured field strength in the location in which the device is used exceed the applicable RF compliance level above the device should be observed to verify normal operation If abnor...

Page 25: ...or the applicator as you may risk exposure to high voltages Never use cartridges and acces sories other than those designated and sold by E Swin SAS for use with ENVIRONMENTAL PROTECTION You must abi...

Page 26: ...ns described in this manual will invalidate the guarantee The company E Swin SAS will not accept liability for any accident arising from failure to observe the instructions contained in this manual or...

Page 27: ...0197 ZA de la Pr v t Rue des C tes d Orval 78550 Houdan FRANCE 01 30 46 37 61 www esw vision com M029GB000F08A rev F 2021 06 11...

Reviews: