E-Swin E-Eye User Manual Download Page 11

20

21

PREPARING THE PATIENT

• Patient has to be settled in an examination and treatment chair lying or sitting.
• The skin must be clean, dry and free of any cosmetic products.  

•  Check that no cream, moisturizing lotion, oil or any other substance has been 

applied to the skin beforehand.

 STAGE 1

F

NOTE:

 

• After 7 seconds of inactivity, 

 screen will switch over to standby mode. Flashing is still possible. Touch the screen to reactivate.

• After around fifteen minutes of inactivity, 

 will switch over to pause mode. Press the 

"play" icon 

13

 

to reactivate.

G

E

6

  TREATMENT OF MEIBOMIAN       BLEPHARITIS USING 

H

445

USING THE ACTIVATION CARD

• Locate the activation card for your 

optic

 cartridge.

•  Insert this card into the slot provided for this purpose, located on the rear of

 

, in 

the base. The location of this slot is easily identifiable by its surrounding sticker.

• Leave this card in place throughout the session.

•  At the end of the session remove the activation card, taking care to avoid damage, and 

replace it in its holder.

• Follow the procedure for turning off the device.

IMPORTANT:

 The activation card for the 

optic

 cartridge must not be damaged, bent or cut. Any damaged card will not be cove

-

red by the E-SWIN guarantee. As a result, any such card will not be replaced and its associated 

optic

 cartridge will be unusable.  

It will then be necessary to purchase a new 

optic

 cartridge, with its activation card. The activation card must not be stored in the 

proximity of a source of heat or electromagnetic radiation.

WHAT YOU NEED TO KNOW:

Each activation card carries the same serial number as the 

optic

 cartridge 

which it accompanies.  Accordingly, each activation card will only work with one 

optic

 cartridge, and must not be used with any other cartridge.

  STAGE 2

•  Upon the initial start-up of 

, "activation codes" (5 characters, followed by 

a further 5 characters) may be requested. Once entered, these codes will not be 

requested a second time.

•  Connect 

 

using the mains lead.

•  The internal test procedure of

 

 will commence, and will last for a few seconds. 

Upon the completion of this phase, screen

 E

 will be displayed automatically.

 STAGE 3

START-UP AND SETTING OF PARAMETERS FOR  

445

445

445

The software version of the device is displayed at the bottom left of the screen and, 

below this, the name of the device and its version number.

•  A personalized 4-digit PIN code may also be entered (see chapter on software options 

in the present manual). 

 STAGE 4

SETTING PARAMETERS FOR  

 DURING THE  

TREATMENT SESSION

•  Select the menu «Treatment of meibomian blepharitis» 

(screen

 F

)

IMPORTANT: 

The «Treatment of meibomian blepharitis» menu can only be accessed 

if your flash credit balance is not zero. 

•  Select the power level 

(screen

 G

)

 

"Selection of energy level" will be displayed at the bottom of the touch screen. This 

screen permits the setting of the required power Level (screen 

G

). The arrows are 

used to increase or decrease the power of the flash delivered, which will then be 

displayed to the left of the power gauge, in joules per centimetre squared (J/cm

2

). 

The power level available for selection ranges from 9 to 13 joules/cm

2

•  Unlocking the security features 

(screen

 H

)

-  Your patient is wearing their 

eye mask

, as a compulsory requirement, 

throughout the session.

-  You are wearing the 

mask

 goggles supplied, for visual comfort during the 

treatment session.

- You have applied 

gel

 to the entire zone to be treated.

-  You are not using

 

 under certain physiological or medical circumstances 

(on pregnant women, on persons taking photo-sensitizing medication, etc.).

-  You have familiarized yourself with the user manual before proceeding with the 

session (specifically the hazard warnings and precautions for use).

-  You have ensured that it is not possible for water to be sprayed onto the

 

.

Once all precautions have been validated, click on the validation icon on the right.

NOTE:

 

 

will not deliver any flash until you have unlocked all the security 

features by clicking on each of the security icon.

Summary of Contents for E-Eye

Page 1: ...User Manual...

Page 2: ...nging and is a practical example of this fact Your patients can now enjoy the benefits of a new treatment for meibomian blepharitis dry eye syn drome using state of the art pulsed light technology wit...

Page 3: ...ETC The manufacturer s guarantee for parts and labour will only be valid if your device is returned in its complete original packaging All items of original packaging in which your was delivered shou...

Page 4: ......

Page 5: ...onal user of must ensure that only persons who fulfil the criteria for use described in the present manual will receive treatment using Before using your the following information should be read caref...

Page 6: ...lement for less than two months On an injured burned or infected skin On skin lesions of unusual appearance which will require investigation On high risk areas blemishes birthmarks warts unhealed woun...

Page 7: ...nection Activation card slot Applicator holding Ventilation grid Visor to protect touchscreen Mobile head Release clips for the mobile head Mobile head Applicator zone Applicator zone Air vents Trigge...

Page 8: ...use of the device 13 Pause icon for the flash treatment session pressing this icon will place in pause mode In this mode no flash can be emitted from the applicator To exit this mode press the same i...

Page 9: ......

Page 10: ...plicator photo 3 Push the extractor into the mobile head to its full length photo 4 Apply pressure simultaneously to the 2 retaining lugs of the mobile head and remove the mobile head photo 5 Remove t...

Page 11: ...nce entered these codes will not be requested a second time Connect using the mains lead The internal test procedure of will commence and will last for a few seconds Upon the completion of this phase...

Page 12: ...n a generous layer minimum 1cm thick on the cheekbone and the tempo ral areas making sure that the gel does not get under the eye mask Locate your mask goggles and put them on If there are other perso...

Page 13: ......

Page 14: ...tivation card and its optic cartridge must be replaced when the number of flashes reaches zero STAGE 2 COOLING OF OPTICAL PARTS During the procedure for the replacement of the optic cartridge before s...

Page 15: ...tridges are not symmetrical so they must be inserted in the right sense and certainly not in the other Check that the cartridge is correctly engaged PRECAUTIONS throughout the procedure for the replac...

Page 16: ...the old activation card and insert the new card supplied with the optic cartridge in the slot provided for this purpose photo I Each activation card carries the same serial number as the optic cartrid...

Page 17: ...on icon 7 The C key can be used to correct the figure in each box SELECT LANGUAGE screen L Click on the flag which corresponds to the required language The Language change confirmed message will be d...

Page 18: ......

Page 19: ...o danger or entail the risk of irreparable damage to your device Once o p t i c cartridge is removed it should be handled by the plastic parts only above all the electrodes the tube or the reflector s...

Page 20: ...sothiazolinone 0 05 0 15 Deionized water Q S 100 For external use only Not to be used on broken burned or infected skin Keep out of reach of children The tool extractor is used for the separation of t...

Page 21: ...of your device Never apply the clean spray directly to the internal components of the applicator APPLICATOR BODY All plastic components of the applicator must be cleaned using a clean cloth which has...

Page 22: ...ols of the optical parts of the applicator to end SECU 03 SECU 04 SECU 05 SECU 07 SECU 09 Your device is blocked Contact our technical department by filling out the contact form on www esw vision com...

Page 23: ...pheric pressure 70 106 kPa MARQUAGE Reference to the treatment of meibomian blepharitis dry eye syndrome using device Unique serial number of your E SWIN Rue des C tes d Orval ZA de la Pr v t 78550 Ho...

Page 24: ...considered If the measured field strength in the location in which the device is used exceed the applicable RF compliance level above the device should be observed to verify normal operation If abnor...

Page 25: ...or the applicator as you may risk exposure to high voltages Never use cartridges and acces sories other than those designated and sold by E Swin SAS for use with ENVIRONMENTAL PROTECTION You must abi...

Page 26: ...ns described in this manual will invalidate the guarantee The company E Swin SAS will not accept liability for any accident arising from failure to observe the instructions contained in this manual or...

Page 27: ...0197 ZA de la Pr v t Rue des C tes d Orval 78550 Houdan FRANCE 01 30 46 37 61 www esw vision com M029GB000F08A rev F 2021 06 11...

Reviews: