TREATMENT OF MEIBOMIAN BLEPHARITIS USING
❙
BEFORE EACH APPLICATION
❙
REMOVAL OF THE MOBILE HEAD
FOR THE CLEANING OF EXTERNAL AND INTERNAL OPTICAL COMPONENTS
• Check the mains lead (to ensure that the latter is not trapped or compressed, and
that no wires are exposed).
• Check the cable connecting the base unit to the applicator (for the absence of
visible deterioration).
• Check the external (photo
1
) and internal (photo
2
) optical components of
.
INSPECTION OF EXTERNAL OPTICAL COMPONENTS
Check that the glass end of the applicator (optical guide) is free of surface defects.
If this is not the case, the mobile head must be replaced. This component can be or
-
dered by visiting
www.esw-vision.com/contact
and choosing the applicable contact
for your country.
INSPECTION OF INTERNAL OPTICAL COMPONENTS
Each time before use and before connecting your
to supply, you must check
the integrity of the internal optical components: the tube, the
filter
(photo
2
) (the
flat component of red-coloured glass) and the optical guide.
• Hold the applicator with the ventilation slots facing you. Insert
the
extractor
between the mobile head and the fixed late
-
ral shell of the applicator (photo
3
).
• Push the
extractor
into the mobile head to its full length
(photo
4
).
• Apply pressure simultaneously to the 2 retaining lugs of the
mobile head, and remove the mobile head (photo
5
).
• Remove the
extractor
and place the applicator carefully
on a flat surface (photo
6
).
• Hold the mobile head, and the
filter
will now be visible
(photo
7
). Apply the
clean
spray to a lint-free cloth (photo
8
), and clean the
filter
using the cloth thus prepared (photo
9
).
• Hold the mobile head, gently move the sidewalls outwards
(photo
10
), carefully remove the
filter
cartridge (photo
11
)
and place it on a flat surface (photo
12
).
• Repeat the same cleaning operation for the internal optical
lens.
The
filter
(the flat component of red-coloured glass) must be
intact and free of cracks, and must show a flat and regular sur
-
face. It must be clean and free of dust. To improve the efficiency
of
, it is recommended that the
clean
spray supplied
with
should be used to clean the
filter
before each use
of the device. Any
filter
which is damaged as a result of incor-
rect maintenance will not be covered by the guarantee (see ins-
tructions for the use of the spray
clean
above).
3
4
5
6
7
10
11
12
1
2
WARNING:
is reserved exclusively for the use of duly authorized ophthalmological professionals.
8
9
You must not use your
if the external and or internal optical components are damaged.