Exchange of electronics / terminal module for Micropilot S, FMR 53X

KA560F/00/a2/05.02
5200 9848
release: 2

Hauser Maulburg

0

1

2

0

1

2

0

1

2

1

6

2

7

3

4

5

11

10

9

8

12

12

ENDRESS

+HA

USE

R

HA

USER

END

RES

S+H

AUS

ER

MIC

ROP

ILOT F

MR

HA

USER

MICR

OPILO

T FMR

IP 6

5

IP 65

Order Code:
Ser

.-No

.:

Order Code:
Ser

.-No

.:

Messbereich

Measur

ing r

ange

Messbereich

Measur

ing r

ange

U 16...36

V DC

4...20 mA

U 16...36

V DC

4...20 mA

max.

20 m

max.

20 m

Made in Ger

man

y             Maulb

urg

Made in Ger

man

y             Maulb

u

rg

T

>70°C :

A

t >85°C

T

>70°C :

A

t >85°C

cover

front panel

bore for
seal wire

electronic
module

view into open

terminal compartment

13

unplug
tool

groove

locked

open

Only identical, original E+H spare parts may be used.
The instrument may only be maintained and repaired by qualified personnel. The device documentation, applicable standards and legal
requirements as well as any certificates have to be observed!

Ex i-devices: The repair has to be performed such, that the voltage resistance of the Ex ia circuits against ground is maintained.

If required, a test can be performed with 500 Veff over a time period of 60 s.

Any repair on an certified device has to be documented!
This includes the device serial number, repair date, type of repair and name of the repair technician.

The custody locking switch can be sealed for custody transfer applications. Access to the electronics can be revented by means of a custody locking switch
that locks the device settings. In this case, after any breaking of the seal (required for exchange of parts), the Micropilot must again be tested and sealed
by an official at commissioning!

• Philips screw driver size 1
• flat screwdriver for M3 / M4
• unplug tool for antenna plug (order no. 5200 7646)
• Allen wrench AF 2.5 / AF 4 / AF 5
• eventually tweezers

Order of disassembly:

The following tools are required for the exchange:

1

2

3

4

5

6

7

8

9

10

11

12

13

Terminal module and electronics are now disassembled
and can be replaced. Assembly is done in reverse order.
Speciall attention has to be payed to the connection of
the antenna cable 13 .
After exchange of electronics, the instrument has to be
enabled for measurement by means of a basic calibration
(see next page).

Unscrew lid from terminal compartment

Disconnect cable at terminal module

Unscrew mounting screws (2 pcs.) at terminal module

and detach ground wire

Pull out terminal module and unplug connectors

on the back side

A new terminal module can now be installed.

Unscrew lid from electronic compartment (4 turns)

If applicable, push up hook and remove display

from holder

Loosen screw at cover and remove cover

Open locking at the front panel and eventually remove seal

Slightly push in both hook at the module housing and ...

... pull off front panel

Unplug supply voltage, signal cable and display

Loosen 2 mounting screws of the electronic module

and pull module out of the housing

Unplug antenna cable with unplug tool

Any changes of the device that affect the product structure (see operating manual) require the installation of an modification type plate (order No. 5200 5627)
with the new order structure. In adition, note these changes in the fields 0D08 and 0D09 of the service matrix (display module VU 331).

Before disassembly, ensure that the operating voltage for the instrument has been switched off.

Summary of Contents for Micropilot S FMR 53 Series

Page 1: ...die Reparatur eines Ger tes zu dokumentieren Hierzu geh rt die Angabe der Ger te Seriennummer Reparaturdatum Art der Reparatur und ausf hrender Techniker 0 1 2 0 1 2 1 6 2 7 3 4 5 11 10 9 8 12 12 ENDR...

Page 2: ...em Reset auf die Werkseinstellungen r ckzusetzen In der Diagnose Ebene im Feld 0A3 die Zahl 333 eingeben Gegen unbeabsichtigtes ndern der eingestellten Ger tedaten sollte die Parametrierung verriegelt...

Page 3: ...Punkt Auswahl Messwert 00 000 12 ton 13 m 14 ft 15 mm 16 inch 17 none Kommunikations adresse 0 15 wenn 0 dann Multidropfunktion Stromausgang auf 4 mA fixiert aktueller Fehler letzter Fehler l sche le...

Page 4: ...hen edge parameter dB cust map valid not active active trig fail count fact map valid not active active module frequen MHz envelope energy dB present edge parameter dB Mapping Edge First echo Tank bot...

Page 5: ...im Bypass verl ngerter Markierungspunkt steht 90 zum Zulauf gilt auch f r Hornantenne 90 Beh lter Bypass Ausrichtung im Schwallrohr Verl ngerter Markierungspunkt in Richtung der Schlitzreihe Beh lter...

Page 6: ...Endress Hauser Maulburg...

Page 7: ...lot must again be tested and sealed by an official at commissioning Philips screw driver size 1 flat screwdriver for M3 M4 unplug tool for antenna plug order no 5200 7646 Allen wrench AF 2 5 AF 4 AF 5...

Page 8: ...nk It is recommended to reset the instrument to the factory settings before the calibration enter 333 in field 0A3 of the diagnostics level The configuration should be locked against unwanted changes...

Page 9: ...0 12 ton 13 m 14 ft 15 mm 16 inch 17 none communication address 0 15 if 0 then multidrop function current output fixed at 4 mA present error previous error clear last error keep erase reset x for rese...

Page 10: ...hen edge parameter dB cust map valid not active active trig fail count fact map valid not active active module frequen MHz envelope energy dB present edge parameter dB Mapping Edge First echo Tank bot...

Page 11: ...gnment in bypass marker point is offset to T s by 90 also applies to horn antenna 90 tank bypass Alignment in stilling well marker points toward slots tank stilling well 90 EN DR ES S HA US ER MI CR O...

Page 12: ...Endress Hauser Maulburg...

Reviews: