23
15
16
20
24
27
21
19
17
22
23
28
5.17 "oil parameters – moisture in oil" screen
The measurand "water content [ppm]" is a calculated
value. For this calculation oil specific parameters (A und B)
are needed. The default factory parameter set is [01]
EEPARAM1 , which contains oil specific parameters for
mineral transfomer oil. The determination of other oil-spe-
cific parameters can be performed by E+E on request.
15 Opens a screen for oil parameter selection (see fig. E).
The navigation is partially possible only by using the con-
trol cross!
16 Oil parameter set - [01] EEPARAM1 is the default factory
setting.
17 Add new oil parameter ser (see fig. G)
18 List of saved oil parameter sets.
A maximum of 10 entries is possible.
19 Select a parameter set (see fig. F).
20 Sends and saves the selected parameter set (under 19) in
the measuring probe.
21 Edit parameter set (see fig. G).
22 Delete parameter set.
23 Cancel action.
24 Enter name for oil parameter set. Navigate via control
cross.
25 Enter A-Parameter (-1999.9...100.0)
26 Enter B-Parameter (0.00..20.00)
27 Confirm and save oil parameters.
28 Opens screen for customer adjustment of the moisture in
oil probe (see fig. H).
A.
E.
f.
G.
18
25
26