30
5.4.3. Reset to factory settings
Customer-adjustment can be reset to the factory settings by checking the appropriate checkbox and subse-
quently clicking the “reset” button.
5.5. Measuring values overview
The tab
measuring values
provides an overview of the retrieved actual measurement values of the flow
meter (transmitter). Clicking on “Fetch values” will retrieve the actual measurement and MIN / MAX values for
flow, volume flow, temperature, mass flow and pressure (only if a pressure transmitter is connected) from the
transmitter – additional the reading of the consumption meter is retrieved as well.
Checking the “Polling” checkbox will retrieve the measuring data from the transmitter at the selected interval.
5.5.1. Reset of the MIN / MAX values
The MIN/ MAX values of each measurand, as stored in the flow meter (transmitter), can be reset by checking
the appropriate checkbox and subsequently clicking the “Clear MIN / MAX” button.
5.5.2. Reset of the consumption counter (totalizer)
The reading of the consumption meter can be reset by clicking the “Reset meter” button.
5.6. Setting up Process Parameters
In the tab
Process Parameters
you can change the Process gas (medium) and set the pressure compensa-
tion
5.6.1. Change the Process Gas
NOTE: This function is only active if the flow meter for a medium different from air has been ordered (see
order code Medium in the data sheet)
Calibration-Gas:
Is the gas (medium) in which the flow meter was calibrated in the factory. Unless otherwise
specified, the flow meter is calibrated at the factory always in air.
Process-Gas:
Is the gas (medium) in the measured process. The adjustable process gases are set at the
factory and can be selected from a list.
The flow meter is factory set to the ordered gas (medium).
If the setting for the process-gas modified or changed between calibration- and process gas, the changed
setting has to be sent to the transmitter. Use “send data to the transmitter and read ...” button.
The “active gas” to which the flow meter is set, you can see in the field basic information.