A too low torque results in a smaller tension force (fixing force)
on the clamping sleeve. There is the risk of injury due to sudden
expulsion of the sensing probe.
A too high torque can lead to permanent deformation
of the clamping sleeve and the sensing probe. This can make
the removal and re-installation difficult or impossible.
REMOVING OF THE PROBE:
1st step:
Firmly hold sensing probe. (Attention: do not bend connection cable)
2nd step:
Slowly loosen the lock nut with a spanner (spanner width 24) until
the expulsion force acts on the probe.
In the installed state, never completely remove the lock nut, only
unscrew it as much as necessary!
3rd step:
After the sensor head has been pulled out of the process up to the
stop, close the ball valve.
4th step:
Probe can now be completely removed.
clamping
sleeve
lock nut
adapter
body
sealing
element 2
sealing
element 1
Pay attention to the sealing element 1 while mounting or removing the probe. This
sealing element has to be put in a proper form.