28
2-point calibration
Humidity
:
Calibration for accurate results over the entire measurement range.
For calibration procedure, see Hardware, chapter 7. “Humidity/Temperature
Calibration”.
1) Place the probe at the reference humidity (lower point).
2) Click on the Humidity 2-Point Calibration button.
(In a separate window, the measured values will appear in both input fields)
3) Replace the value in the input field "Humidity Reading" with the reference
humidity. (Value of the saline solution or display of HUMOR 20)
4) By clicking on "Save", the humidity reading of the transmitter will be
adjusted to the reference humidity. Now the 30-minute stabilisation
period starts.
5) Place the probe at the reference humidity (high point).
6) Before continuing wait till the 30-minute stabilisation period is over.
7) Replace the value in the input field "Humidity Reading" with the reference
humidity. (Value of the saline solution or display of HUMOR 20)
8) By clicking on "Save", the humidity reading of the transmitter will be
adjusted to the reference humidity.
9) The process is complete when the message "Two-point calibration
successful" appears.
1-point calibration
Temperature
: If the working range is limited to a narrow temperature range, one-point calibration
will be sufficient within this working range.
1) Place the probe at the reference temperature and allow stabilisation for
approx. 30 minutes.
2) Click on the Temperature 1-Point Calibration button.
The measured value will appear in both input fields. (see additional
window)
3) Replace the value in the input field "Temperature Reading" with the
reference temperature.
4) By clicking on "Save", the temperature reading of the transmitter will
be adjusted to the reference temperature.
5) The process is complete when the message "Calibration Successful"
appears.