19
Self diagnostics via LEDs on the circuit board:
•
Green LED
flashing
⇒
Supply voltage applied / Microprocessor is active
•
Red LED
constantly lit
⇒
Humidity sensor element damaged
flashing
⇒
Humidity sensor element accruing moisture (condensation!)
Self diagnostics via display (where available):
Error 1
⇒
Humidity sensor element damaged
Error 2
⇒
Humidity sensor element moistened (condensation!)
Error 3
⇒
Temperature sensor element damaged
Error 4
⇒
Temperature input short circuit
Definitions:
•
Error
possible cause
⇒
Measures / Help
•
Display shows incorrect values
Error during re-adjustment of the transmitter
⇒
Reset to factory calibration and repeat the calibration routine
Filter soiled
⇒
Replace filter
Sensor defective
⇒
Replace sensor
Output configured incorrectly
⇒
PC - Software
•
Transmitter failure
no supply voltage
⇒
Check wiring and supply voltage
⇒
only green LED is illuminated continuously
⇒
Electronics defect
⇒
contact the manufacturer
8.6
Replacement of sealing element
Because of repeated installations and various other circumstances the sealing
element can get damaged. The customer can do a replacement of this sealing
element (refer to drawing).
8.5
Self diagnostics and error messages
9.
REPLACEMENT PARTS / ACCESSORIES
Description
Order Code
- Stainless steel filter
HA010110
- Display and housing cover in metal
D05M
- Display and housing cover in polycarbonate
D05P
- Replacement probe
- for EE36 with 2m cable
HA010902
- for EE36 with 5m cable
HA010905
- for EE36 with 10m cable
HA010910
- Replacement sensor
- Humidity sensor with sensor data
FE10
- Temperature sensor with sensor data
TE38
- RS232 interface cable, 2m
HA010301
- Mounting rail bracket
HA010203
- Ball valve set
HA050101
- Sealing element
HA050309
Hardware
sealing element