Operating manual EE211 Heated humidity and temperature sensor

5.2.2  Address Switch

Address Switch

Address Switch

Address setting via EE-PCS Product Configuration Software

All DIP switches at position 0 → address has to be set via

 

Product Configuration Software

Factory setting: 239 (permitted values: 1…247).

Example: Slave address is set via configuration software.

Address setting via DIP switch

Setting the DIP switches to any other address than 0, overrules the  

slave address set via configuration software (permitted values: 1…247).

Example: Slave address set to 11 (= 0000 1011 binary).

0 0 0 0 0 0 0 0

0
1

1 1 0 1 0 0 0 0

0
1

5.2.3  Digital Settings

Factory settings

Selectable values

Baud rate

9600

9600, 19200, 38400, 57600, 76800, 115200

Data bits

8

8

Parity

EVEN

None, odd, even

Stop bits

1

1 or 2

Slave address

239

1...247

Tab. 2 

Data transmission

ID address, baud rate, parity and stop bits can be set via:

1.  EE-PCS, Product Configuration Software and the appropriate configuration cable.

2.  Modbus protocol in the register 60001 (0x00) and 60002 (0x01). See Application Note Modbus 

AN0103 (available on www.epluse.com/EE211)

The measured values are saved as a 32 bit float value and as 16 Bit signed integer, see the Modbus 

Register map below.

The serial number as ASCII-code is located at read register address 30001-30008 (16 bit per address).

The firmware version is located at register address 30009 (bit 15...8 = major release; bit 7...0 = minor 

release).

The sensor name is located at register address 30010.

5.2.4  Modbus Register Map

FLOAT 32 bit (read register):

Function code /  

Register number

1)

 [Dec]

Register address

2)

 [Hex]

Parameter name

Unit

3)

30026

0x19

temperature

[°C], [°F]

30028

0x1B

relative humidity 

[%]

30030

0x1D

water vapour partial pressure

[mbar], [psi]

30032

0x1F

dew point temperature

[°C], [°F]

30034

0x21

wet bulb temperature

[°C], [°F]

30036

0x23

absolute humidity

[g/m³], [g/ft³]

30038

0x25

mixing ratio

[g/kg], [gr/lb]

30040

0x27

specific enthalpy

[kJ/kg], [BTU/lb]

30042

0x29

frost point temperature

[°C], [°F]

Tab. 3 

Float (read register)

Summary of Contents for EE211

Page 1: ...BA_EE211_e V1 3 technical data are subject to change Operation Manual EE211 HEATED HUMIDITY TEMPERATURE SENSOR...

Page 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Page 3: ...7 3 6 1 Examples 7 4 Assembly and Installation 7 5 Setup 8 5 1 EE PCS Product Configuration Software 8 5 2 Digital Interface RS485 with Modbus RTU Protocol 8 5 2 1 Hardware Bus Termination 8 5 2 2 Add...

Page 4: ...ed due to failure to comply with the applicable regulations operating instructions or the operating conditions Consequential damages are excluded from the liability 1 1 Explanation of Symbols This sym...

Page 5: ...hem clearly as faulty A faulty device may only be investigated and possibly repaired by qualified trained and authorized staff If the fault cannot be fixed the device shall be removed from the system...

Page 6: ...ensing head E E proprietary coating of the sensing element and sealed solder pads minimize the impact of corrosive agents Maximum humidity of 76 RH at the sensor eliminates the drift caused by exposur...

Page 7: ...using the EE PCS Product Configuration Software and the optio nal USB configuration adapter HA011066 3 6 1 Examples Factory setup voltage output 0 5 V or 0 10 V corresponds to 0 100 RH After switching...

Page 8: ...the USB configura tion adapter HA011066 PC HA011066 Fig 3 Configuration adapter 5 1 EE PCS Product Configuration Software 1 Download the EE PCS Product Configuration Software from www epluse com confi...

Page 9: ...set via 1 EE PCS Product Configuration Software and the appropriate configuration cable 2 Modbus protocol in the register 60001 0x00 and 60002 0x01 See Application Note Modbus AN0103 available on www...

Page 10: ...ter number1 Dec Register address2 Hex Parameter name 60001 0x00 Slave ID5 modbus address 60002 0x01 Modbus protocol settings6 Tab 5 Integer 16 bit read and write register INFO read register Function c...

Page 11: ...dress Hi Starting address Lo No of register Hi No of register Lo CRC Request Hex EF 03 00 19 00 02 EB 7B Response Hex EF 03 04 38 5F 41 D5 0A E3 Modbus ID address Function code Byte count Register 1 v...

Page 12: ...roves the measurement accuracy of a device The specimen is compared with the reference and brought in line with it An adjustment can be followed by a calibration which documents the accuracy of the ad...

Page 13: ...e a Windows PC using the EE PCA product configuration adapter and the cable HA011057 see data sheet EE PCA EE PCS free download from www epluse com configurator shall be installed on the PC Start EE P...

Page 14: ...lsoure material Polycarbonate UL94V 0 with Display UL94HB approved T probe material Stainless steel 1 4571 Protection class IP65 NEMA 4 Cable gland M20 x 1 5 Electromagnetic compatibility EN61326 1 In...

Page 15: ...15 Operating manual EE211 Heated humidity and temperature sensor...

Page 16: ...0 6460 Office Stuttgart Tel 49 151 538 37 500 E E Elektronik Italy info epluse it Tel 39 02 2707 86 36 E E Elektronik France info epluse fr Tel 33 4 74 72 35 82 E E Elektronik USA office epluse com Of...

Reviews: