8
Operating manual EE211 Heated humidity and temperature sensor
5 Setup
The EE211 is ready to use and does not require any configuration by the user. The factory setup of
EE211 corresponds to the type number ordered. For ordering guide please see data sheet at www.eplu-
se.com/ee211. If needed, the user can change the factory setup by using the USB configuration adapter
(cod HA011066) and the EE-PCS, Product Configuration Software.
The user can change the scaling and the analogue output signal, the digital settings and perform RH
and T adjustment.
Note:
The EE211 may not be connected to any additional power supply when using the USB configura-
tion adapter HA011066.
PC
HA011066
Fig. 3
Configuration adapter
5.1 EE-PCS Product Configuration Software
1. Download the EE-PCS Product Configuration Software from
www.epluse.com/configurator
and
install it on the PC.
2. Connect the E+E device to the PC using the USB configuration cable.
3. Start the EE-PCS software.
4. Follow the instructions on the EE-PCS opening page for scanning the ports and identifying the
connected device.
5. Click on the desired setup or adjustment mode from the main EE-PCS menu on the left and follow
the online instructions of the EE-PCS
5.2 Digital Interface RS485 with Modbus RTU Protocol
5.2.1 Hardware Bus Termination
If required, the bus termination shall be realized with 120 Ohm resistor, jumper on the board.
• Jumper mounted
= bus terminated
• Jumper not mounted = bus not terminated