Operating manual EE211 Heated humidity and temperature sensor

Outstanding long term stability under high humidity conditions

The operation principle of EE211 copes with the causes for poor long-term stability of non-heated  

sensors at continuously high humidity. The constant over-temperature of the EE211 sensing head 

(approx. 5 °C = 9 °F) means max. 76 % RH humidity at the sensors and enables following benefits:

•  The sensing head of EE211 stays dry even under condensing conditions, which prevents dust and dirt 

from sticking to the sensor and leads to outstanding long-term stability.

•  The combination of dry sensing head, E+E proprietary coating of the sensing element and sealed 

solder pads minimize the impact of corrosive agents.

•  Maximum humidity of 76 % RH at the sensor eliminates the drift caused by exposure to continuous 

high humidity.

3.2  Dimensions in mm (inch)

80.6

(3.17“)

101

(3.98“)

Ø 12

(Ø 0.47“)

~ 55

(~2.16“)

5

(0.2“)

90 

± 0,3

(3.54

 

+0.11

)

46

(1.81“)

CABLE GLAND 

M20

Ø 13

(Ø 0.51“)

34

(1.34“)

Basis Unit:

Temperature Probe:

M12x1

50 

(2”)

8

(0.3”)

Ø6 

(

0.24“)

Ø1

2

(0.47”)

75 

(2.9”)

Fig. 1 

Dimensions of basis unit and temperature probe

3.3  Connection Diagram

Modbus RTU

15...35 V DC 

 

24 V AC ± 20%

OUT1

OUT2

V

mA

V

mA

15...35 V DC 

 

24 V AC ± 20%

Output:   0...5 V  

 

 0...10 V

 

 0...20 mA

 

 4...20 mA

Output:

EE211-M1J3 - Digital Output

EE211-M1A2/3/5/6 - Analogue Output

Bus termination resistor 120 Ω (jumper)

Address switches

Selection output signal U / I

Setup interface

Fig. 2 

Connections and setup

Summary of Contents for EE211

Page 1: ...BA_EE211_e V1 3 technical data are subject to change Operation Manual EE211 HEATED HUMIDITY TEMPERATURE SENSOR...

Page 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Page 3: ...7 3 6 1 Examples 7 4 Assembly and Installation 7 5 Setup 8 5 1 EE PCS Product Configuration Software 8 5 2 Digital Interface RS485 with Modbus RTU Protocol 8 5 2 1 Hardware Bus Termination 8 5 2 2 Add...

Page 4: ...ed due to failure to comply with the applicable regulations operating instructions or the operating conditions Consequential damages are excluded from the liability 1 1 Explanation of Symbols This sym...

Page 5: ...hem clearly as faulty A faulty device may only be investigated and possibly repaired by qualified trained and authorized staff If the fault cannot be fixed the device shall be removed from the system...

Page 6: ...ensing head E E proprietary coating of the sensing element and sealed solder pads minimize the impact of corrosive agents Maximum humidity of 76 RH at the sensor eliminates the drift caused by exposur...

Page 7: ...using the EE PCS Product Configuration Software and the optio nal USB configuration adapter HA011066 3 6 1 Examples Factory setup voltage output 0 5 V or 0 10 V corresponds to 0 100 RH After switching...

Page 8: ...the USB configura tion adapter HA011066 PC HA011066 Fig 3 Configuration adapter 5 1 EE PCS Product Configuration Software 1 Download the EE PCS Product Configuration Software from www epluse com confi...

Page 9: ...set via 1 EE PCS Product Configuration Software and the appropriate configuration cable 2 Modbus protocol in the register 60001 0x00 and 60002 0x01 See Application Note Modbus AN0103 available on www...

Page 10: ...ter number1 Dec Register address2 Hex Parameter name 60001 0x00 Slave ID5 modbus address 60002 0x01 Modbus protocol settings6 Tab 5 Integer 16 bit read and write register INFO read register Function c...

Page 11: ...dress Hi Starting address Lo No of register Hi No of register Lo CRC Request Hex EF 03 00 19 00 02 EB 7B Response Hex EF 03 04 38 5F 41 D5 0A E3 Modbus ID address Function code Byte count Register 1 v...

Page 12: ...roves the measurement accuracy of a device The specimen is compared with the reference and brought in line with it An adjustment can be followed by a calibration which documents the accuracy of the ad...

Page 13: ...e a Windows PC using the EE PCA product configuration adapter and the cable HA011057 see data sheet EE PCA EE PCS free download from www epluse com configurator shall be installed on the PC Start EE P...

Page 14: ...lsoure material Polycarbonate UL94V 0 with Display UL94HB approved T probe material Stainless steel 1 4571 Protection class IP65 NEMA 4 Cable gland M20 x 1 5 Electromagnetic compatibility EN61326 1 In...

Page 15: ...15 Operating manual EE211 Heated humidity and temperature sensor...

Page 16: ...0 6460 Office Stuttgart Tel 49 151 538 37 500 E E Elektronik Italy info epluse it Tel 39 02 2707 86 36 E E Elektronik France info epluse fr Tel 33 4 74 72 35 82 E E Elektronik USA office epluse com Of...

Reviews: