DYSIS Ultra 2.0 Digital Colposcope Instructions for Use
0330-53095 R4
Revision Date 25-Feb-2021
Page
52
of
61
20.9.2.
Software Messages
Message
Possible Cause/User Action
15 Spray cycles have completed. Check
acetic acid level
Make sure that there is enough acetic acid in the reservoir. If not,
please refill the container.
All visit data/images will be discarded.
Press OK to confirm
A message shown when CANCEL is pressed after a direct DYSIS
exam. By selecting OK, all data from the visit will be discarded.
An error has been detected in the
camera/syringe. Please contact DYSIS
When this message is displayed, DYSIS Exam is disabled. Shut
down the DYSIS Ultra 2.0 Digital Colposcope and then start it up
again. If message persists, please contact DYSIS Medical service
department.
An error has been detected: Code: XXXX
Press OK to shutdown
Please contact the Administrator to export logs and forward the
logs to DYSIS Medical service department for investigation.
Are you sure you want to delete image?
By selecting OK, the selected image will be deleted.
Due to excessive movements, no map will
be calculated
There were too many micro-movements, causing failure in image
alignment and inability to calculate the map
Failed to export report. Can't access
export directory
Please check the export directory is valid and accessible through
an Administrator account.
Failed to export report. Please conFig an
export path in settings
Please check the export directory is valid and accessible through
an Administrator account.
It has been xx days since the last
calibration. Please calibrate now
Weekly calibration is required. This message will be displayed to
perform the calibration. No mapping will be available until DYSIS
Ultra 2.0 Digital Colposcope has been calibrated.
It has been 30 days since the acetic acid
consumables were changed. Please
change now
Please change the DYSIS acetic acid kit.
No colposcopy directed biopsy sites
indicated. Confirm?
This message is displayed when Show Map is selected without
having added any biopsy points. By selecting OK, you confirm no
biopsy points need to be added and the map will be displayed.
By pressing cancel, you have the option to add biopsy markers.
No initial prediction selected. Please
select prediction to continue
Please select the Initial Predication from the right-side menu for
the colposcopy examination.
No final prediction selected. Please select
Prediction to continue
Please select a Final Prediction from the right-side menu for the
colposcopy examination.
No map will be calculated. Do you still
wish to exit?
This message is displayed when STOP is selected during the
mapping procedure. No map will be calculated since not enough
images were recorded. For the map to be calculated, please
press NO.