DYSIS Ultra 2.0 Digital Colposcope Instructions for Use
0330-53095 R4
Revision Date 25-Feb-2021
Page
45
of
61
Fig 45: Calibrate option
If calibration has expired, a message will appear to prompt the user to calibrate the system (Fig 46).
Acknowledge the message by selecting OK and access the CALIBRATE option from the left panel of the HOME
screen.
Fig 46: Calibration message
For the calibration procedure, the user should place the provided grey card on its mount (jig) and then connect it
to the speculum connector (Fig 47).