When
is displayed, the two visits can be reviewed independently. When this symbol is aqua and
unlinked, both sides can be manually played and images paused or advanced separately. Each visit is controlled
by a set of buttons below each image and the map histogram is displayed respectively below each visit.
When the chain links are locked
the images will play, move and magnify simultaneously,
controlled by a single set of buttons at the bottom of the screen.
The buttons below each image allow images to be advanced individually frame by frame (taken at 7 second
intervals), at
normal speed and
normal speed, by selecting the appropriate option.
Using
and
buttons, the images can be advanced or reversed individually. The time point of each image
is displayed on the right side of each screen.
Switching between Captured and Aligned images (Fig 43) allows to view the raw images as captured (best for
details), or the images aligned by the software for the map calculation. The images need to be aligned in order to
view the DYSISmap.
When viewing the aligned images, two additional buttons become available:
.
By selecting
, the map will be displayed or hidden over the reference image. By selecting
, the biopsy
annotations will be displayed and hidden over the reference image.
Fig 43: SMARTtrack with comparison displayed
To change the exams for comparison, select a visit from the box below each image and select the exam to
review. They will be uploaded into the viewing area.