DYSIS Ultra 2.0 Digital Colposcope Instructions for Use
0330-53095 R4
Revision Date 25-Feb-2021
Page
44
of
61
19.
Additional Functions
In the
menu, there are additional features and settings available to the Administrator and the common user
based upon the role assigned.
19.1.
Simulator
The simulator contains 5 patient cases which allows the clinician to use the software on full patient cases prior to
use on a patient. This may be used for training on the DYSIS Ultra 2.0 Colposcope.
To start the simulation case, select Simulator then Enter DYSIS Mode to access the functionality of the user
interface and perform a patient exam. Use all of the exam function and perform a DYSIS exam with the simulated
exam. Refer to Section 16.4 on how to perform a DYSIS Exam.
19.2.
Purge syringe
The purge syringe option is used to expel any air in the tubing. When this is selected, the acetic acid applicator
will activate and spray acetic acid from the diffusor located on the speculum connector. To minimize spray, place
a paper towel in front of the diffusor. Press Purge until the air has been expelled from the tubing.
19.3.
Settings
The following are available under Settings.
19.3.1.
Info
The INFO tab displays the software version in use, for example 2.2. In addition, serial numbers of components,
such as the camera, are displayed. It is important to know this information when contacting the DYSIS service
and maintenance team.
19.3.2.
Password
The PASSWORD function allows users to manage their own login and passwords. Usernames and passwords are
case sensitive, and passwords need to be at least five characters. Please enter information in all fields shown,
LOG IN NAME, REAL NAME and PASSWORD to create new users. Notice that REAL NAME will be included in
patient reports.
19.4.
Calibrate
The calibration function can be accessed through the left panel of the HOME screen (Fig 45).