UPR700 Microprocessor-Based Pressure/Process Indicator
45
Fig. 14 UPR700 Board Location
9.2.2 G
ENERAL
C
ALIBRATION
P
ROCEDURE
1. Set the dip switches to calibration mode as shown above. The upper display should show
CAL
while the lower display shows
UPR
.
2. Use the
▼
or
▲
keys to show the following functions:
•
Firmware revision
•
Zero input counts (ZERO)
•
Pressure input counts (STR)
•
Reference junction counts (RJ)
•
Remote set point, linear temperature input and line resistance for RTD input (RSP.RL)
•
Thermocouple and RTD input (TC.RTD)
•
Digital inputs status (DIG.IN)
•
Maximum Power Consumption
•
Minimum Power Consumption
3. The display values for analog inputs are scaled from 0 to 25,000 counts; it is also linear for RTD
input.
4. Use the
▼
or
▲
keys to select a display value from 0 to 10 and to check the linearity of output
circuit at 0%, 10%, 90% and 100%.