hIgh VolT InsTruCTIons
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017
44
Detailed High Volt Setup
1. Press the HI VOLT key.
The High Volt channel and the default settings for High Voltage electrode pulsed stimulation are automatically
selected. Using the dedicated High Volt (HV) channel attach leads and place electrodes on the patient now. For a
High Volt Quick Setup
1. Press HI VOLT.
Electrode Treatment
• Plug in lead wire to the HIGH VOLT OUTPUT JACK (HV).
• Attach electrodes to patient.
Probe Treatment
• Press CHANNEL TOGGLE key until the PROBE LED is lighted.
• Plug the optional MultiStim probe into the STIM OUTPUT JACK.
• Attach dispersive electrode.
2. Choose the POLARITY. Bipolar is not available for probe treatments. Note: Polarity must be selected
before customizing a treatment.
3. TIME. Select TIME by using the TIME ARROW keys for Electrode Pads treatments only. Probe Treatment
TIME counts up.
4. Select (or setup) a PULSE RATE RANGE (High/Low) or set a single pulse rate by pressing the MODE toggle
key and selecting RATES. Use the arrow keys next to RATE-START and RATE-STOP to set the Pulse Rate
Range.
5. Set CONTRACTION/REST times by pressing the FUNC (Function key). Use the CONT/REST toggle key
to make your selection.
6. RAMP TIME. While in FUNCTION mode use the RAMP TOGGLE key to select RAMP TIME.
7. Raise the INTENSITY to the desired level. For a probe treatment, make sure probe is in contact with
patient’s skin when raising intensity.
8. Press START. Treatment time will begin to count-down for an Electrodes treatment. For a probe treatment,
press ON/OFF on the probe handle after pressing the device START key to activate the treatment. Time will
begin counting up.
9. STOP. Press and hold the FUNCTION and press the STOP key to stop only the focus treatment appearing
in the Treatment Screen. Pressing STOP alone, stops all treatments operating on the device. Using the
TIME arrow keys to bring treatment time to zero will also stop a treatment.