Dynatron
®
150
plus
Enter Soundhead Parameters
39
b. Select 50% Duty Cycle: With the 50% LED lighted, enter the value for
Z – 2 MHz.
c. Select 20% Duty Cycle: With the CONT LED lighted, enter the value for
Temp – 2 MHz.
d. Select 10% Duty Cycle: With the 10% LED lighted, enter “0” (zero) as the
value for Coupling. The coupling feature is not available with old-style
soundheads.
6. Enter 3 MHz Values. Press the FREQ key to select 3 MHz. Make sure the 3 MHz
LED is lighted. Locate the 3 MHz column on the printout, and enter the numbers
as follows:
a. Select CONT Duty Cycle: The CONT and 3 MHz LEDs are lighted. Enter
the value for F1 – 3 MHz by pressing the time selection keys until the
desired value is displayed in the time display.
b. Select 50% Duty Cycle: With the 50% LED lighted, enter the value for
Z – 3 MHz.
c. Select 20% Duty Cycle: With the CONT LED lighted, enter the value for
Temp – 3 MHz.
d. Select 10% Duty Cycle: With the 10% LED lighted, enter “0” (zero) as the
value for Coupling. The coupling feature is not available with old-style
soundheads.
7. Store New Parameters. After you have entered all parameters, press START to
store them in the device’s memory. Then press STOP to exit this mode.
The above procedure must be performed for each separate soundhead for the device.
Turn the device off before attaching the next soundhead, then turn the device on again
with the soundhead firmly plugged in.
Summary of Contents for 150plus
Page 1: ...SERVICE MANUAL Dynatron 150plus ...
Page 6: ...Dynatron 150plus vi ...
Page 20: ...Dynatron 150plus 14 ...
Page 24: ...Dynatron 150plus 18 ...
Page 32: ...Dynatron 150plus 26 ...
Page 42: ...Dynatron 150plus Beam Profile 36 2 cm2 Head Near Field 1 MHz 3 MHz ...
Page 46: ...Dynatron 150plus 40 ...
Page 50: ...Dynatron 150plus 44 ...
Page 54: ...Dynatron 150plus 48 ...
Page 59: ...53 ...
Page 60: ...54 ...
Page 61: ...55 ...
Page 62: ...56 ...
Page 63: ...57 ...
Page 64: ...58 ...
Page 65: ...59 ...