![Dynamica Velocity 15HR Instruction Manual Download Page 52](http://html.mh-extra.com/html/dynamica/velocity-15hr/velocity-15hr_instruction-manual_2548262052.webp)
・
When using tubes on the market, perform operation under the allowable speed or the allowable RCF
specified by the manufacturer. Otherwise the tubes may be broken during operation.
・
Precautions in using centrifugal filter unit when using the FAS15E/FAS15F rotor. 1
)
Be sure to cap the tubes before the centrifugation when the tubes can be capped.
2
)
Load the tubes to the rotor not to stick the tube caps out of the rotor when the tubes can not be capped.
Spare parts
Lubricant for O-rings or the threaded portion of the rotor is not included in this rotor. Please order applicable
ones selecting from the following list (designate the part numbers).
O-rings are consumable parts. We recommend that O-rings are replaced with new ones periodically.
For FAS15E/FAS15F/FAS15G rotor
Part No. 483719
Silicone grease (vacuum grease)
Part No. 84810601 Lubricant for screw
For FAS15E rotor
O-ring for the rotor cover
Part No. 84521015
O-ring for the cover-locking knob (5pcs./set)
For FAS15F rotor
Part No. 84521015
O-ring for the cover-locking knob (5pcs./set)
Part No. 4348812
O-ring for the rotor cover (1pc./set)
For FAS15G rotor
Part No. 84521015
O-ring for the cover-locking knob (5pcs./set)
Part No. 84522007
O-ring for the rotor cover (1pc./set)
!
CAUTION
:
When using tubes outside 4-25
℃
,
tubes may be broken or deformed during operation.
Test them before actual operation and perform actual operation after checking that
there is no inconveniences such as the above.
Filter
Micro tube
Tube cap
!
CAUTION
:
When using the marketed centrifugal filter unit,
load the tubes to the rotor not to stick the tube
caps out of the rotor if the tubes can not be
capped.
Otherwise the rotor may abnormally vibrate
depending on the tube shape during the
centrifugation.
Example of incorrect setting
O-ring for the cover-locking knob
Example of centrifugal filter unit
Load the tubes to the rotor
not to stick the tube caps
out of the rotor when the
tubes can not be capped.
Be sure to cap the tubes
before the centrifugation.
Example of correct setting
Example of correct setting