![Dynamica Velocity 15HR Instruction Manual Download Page 37](http://html.mh-extra.com/html/dynamica/velocity-15hr/velocity-15hr_instruction-manual_2548262037.webp)
6-5
6. Troubleshooting
(3) Insert the attached tool for setting the rotor horizontally into the hole for unlocking
the door to fit the tool for setting rotor into the shaft inside the centrifuge.
(4) By turning the tool for setting the rotor at 180°, the door is unlocked automatically in
the centrifuge and the door is opened to a few cm automatically.
■
After the door is opened, remove the tool for setting the rotor. Then be sure to
position the plug into the hole for unlocking the door. It is not necessary to place
the shaft to the original position.
(5) Open the door and take out the samples.
WARNING
:
●
When you first turn on the POWER switch after you unlock the door as described
above, the door lock might work automatically to adjust it’s position. Do not insert
your fingers or an object into the door hook hole. Otherwise your fingers might be
injured or the centrifuge might be damaged.
●
Be sure to remove the tool for setting the rotor from the centrifuge if it is not
necessary to inset it to the centrifuge. It is very dangerous because the tool
contacting the shaft rotates if the tool is being inserted into the centrifuge when you
close the centrifuge door after turning on the POWER switch.
!
NOTE
:
●
You can not see the shaft for unlocking the door, because it is inside the centrifuge. Insert
the attached tool for setting the rotor horizontally until the tool contacts the shaft. Then fit
the tool for setting rotor into the shaft for unlocking the door by turning the tool clockwise or
counterclockwise slowly.
Tool for setting the rotor
Window of the door