Installation and Testing
5.1
Mounting
All Dimensions in mm
The G91 can be mounted on either side of the wheelchair to accommodate users
preference.
G91 must be mounted using both M5 fixing screws. The selected screw is to provide at
least 4mm, best 6mm thread engagement and must not allow penetration of more
than 7mm into the case. Maximum tightening torque of the screws is 2Nm.
If the G91 is tray mounted, particular care must be taken to prevent direct collision of
the product into immovable objects, such as a desk.
Water ingress should be prevented by putting a seal between the G91 and the tray. If
the G91 is designed to rotate then the housing should be designed to cover and seal
the cables entering the G91.
A plastic insert is available to cover the programming socket and is recommended in
a tray mount. RS part no. 225-899
Summary of Contents for DX-REMG91
Page 1: ......
Page 2: ......
Page 6: ...GBK64048 Issue 3 February 2012 ...
Page 8: ...GBK63648 Issue 3 February 2012 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 10: ...GBK63648 Issue 3 February 2012 4 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 33: ...Chapter 3 G91 Operation 27 3 9 1 Joystick Only Menu Map ...
Page 40: ...GBK63648 Issue 3 February 2012 34 Pattern 3 1 2 1 3 1 4 Pattern 4 1 5 1 6 1 2 1 3 1 4 ...
Page 45: ...Chapter 4 G91S Operation 39 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 51: ...Chapter 5 Installation and Testing 45 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 61: ...Chapter 6 Programming G91 55 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 69: ...Chapter 9 Appendices 63 8 3 External Dimensions All dimensions in mm ...
Page 79: ...Chapter 9 Appendices 73 ...