4. Make sure to completely bleed out the pressure inside HPFM CAT, and confirm that with pressure gauge reading.
5. Click
OK
to start set-zero (See the screen above.) Once hearing a beep, start increasing the pressure
inside HPFM CAT by turning the air valve to
Air Flow In
position.
6. Stop pressuring at 550 kPa by turning the air valve to
Closed
position.
7. If there is any unsatisfactory data collected, you can check
Set Ignore Points
box and high light those
points (shown in dark color.) You can also remove these highlighted points a, b, c, d, by clicking on
DEL
button.
Normally a 1st order regression will remove dP offsets across the range of a transient data sets.
a. To Ignore points check the
Set Ignore Points
Box in the bottom left hand corner, and then
select the points you want to ignore.
b. To delete points, select the points you want to delete and click
DEL
.
c. To retrieve deleted points click on the
UNDO
button in the bottom right hand corner.
d. To undo the regression select the
UNDO
button in the bottom left hand corner
8. You may increase
Order
to 2 and click on
REG
button to perform a regression. A new window will pop up to
indicate the regression result. Click
Yes
to save it.
9. Saved set-zero results applied to all subsequent transient dP calculations. Removing drift or temperature
changes to the pressure sensor readings. Set-zero regressions can be seen in the data viewer.
Summary of Contents for HPFM3
Page 49: ...Revision JUNE 2017 49 b Click on the STATIC OFFSET set the new value to zero and SAVE ...
Page 73: ...Revision JUNE 2017 73 Red Range ...
Page 74: ...Revision JUNE 2017 74 Yellow Range ...
Page 75: ...Revision JUNE 2017 75 Grey Range ...
Page 76: ...Revision JUNE 2017 76 Brown Range ...
Page 77: ...Revision JUNE 2017 77 Zero Adjustment Measurement ...