BlueWave® QX4® User Guide
21
Irradiation
Once all configurations are ready, navigate to the RUN row (Figure 23) using the up and
down buttons. Press the center run button to start irradiation in all LED Heads.
Figure 23. Select RUN Option
Figure 24. Irradiation Screen
During irradiation, the steps and cycles will count down to indicate the time remaining
on the current curing session. Pressing the center run button during the irradiation
cycle will stop the irradiation. The Footswitch can also be used instead of pressing the
center run button.
An Indicator Light will show on the bottom of the front panel to show which LED Heads
are active during the irradiation cycle. A green light shows that the LED Head is active
with no faults. An orange indicator light shows that the corresponding LED Head has
reached the user-set temperature-warning setting. The LED Head will continue to
function as normal until it reaches a critical temperature condition.