1 6
U S E R G U I D E
|
B L U E W A V E
®
Q X 4
®
System Setup
System Connections
Power Cable Connection
—
Attach the Power Cord to the Power Cord Receptacle
located on the
unit’s
back panel (Figure 4). Press the Power Cord firmly into the
receptacle until it clicks and locks into place. The unit is now powered and ready to be
turned on with the On/Off Switch.
To remove the Power Cord, pull the retaining body of the Connector to unlock it from
the receptacle.
LED Head Connection Jacks
—
Along the right of the
Controller’s Rear Panel,
there
are four LED Head Connection Jacks labeled CH 1 - 4. The Connectors are keyed so
they may require slight rotation to align with the keying elements of the connector pair.
Once the keyways are aligned, press the LED Head Connector into the Jack until it
clicks and locks in place.
NOTE: DO NOT rotate the Connectors once installed, they are not threaded, and
damage may occur.
To remove the LED Head, grasp the metal Outer Retaining Ring Body of the Connector
and pull away from the Controller to unlock it from the Jack.
Footswitch Connection
(Optional)
—
Located on the
middle of the Controller’s
rear
panel. It can be used as an optional irradiation trigger.
PLC Connection Terminals
—
There are input and output PLC Connection Terminals
that can be used to integrate the unit to an automated assembly line. See the
PLC
Operation Section
for more details.
A low signal (0V) input on PLC_ENABLE switches the QX4 to PLC mode. In PLC mode,
the Front Control Display displays the PLC connection and locks out the screen input.
A high signal (24V) input on MASTER INTERLOCK locks out all the channel output.