Table 9. Alarms Chart for BlueWave® MX-MIM
Alarm # Alarm/Warning Description
Cause/Corrective Action
1
MX-150 Lightguide Alarm
Verify lightguide is installed correctly and fully
seated.
2
Emitter Interconnect Cable
Verify emitter interconnect cable installation.
Verify minimum clearance (5 inches) is not
exceeded.
Check for bent pins on all D-SUB connections.
Replace emitter interconnect cable.
4
Controller Temperature
Alarm >78°C
Reduce obstructions from controller airflow.
Check air flow and look for fan obstructions.
Check vent min clearance distance (1 inch).
Reduce ambient temperature of controller.
5
Emitter LED
Return emitter to Dymax.
6
Emitter Firmware
Return emitter to Dymax.
7
Emitter Lost Comms
Verify emitter interconnect cable installation.
Verify minimum clearance (5 inches) is not
exceeded.
Replace emitter interconnect cable.
8
Emitter Interlock is open
Verify interlock jumper or external control has not
been opened/defeated.
Verify PLC command structure.
9-15
Internal Failure
Return emitter to Dymax.
16
Controller Fan Alarm
Return controller to Dymax.
17
Controller Temperature
Warning >69°C
Reduce obstructions from controller airflow.
Check air flow and look for fan obstructions.
Check vent min clearance distance (1 inch).
Reduce ambient temperature of controller.
18
Controller Slow Fan Warning
Check 24 VDC supply is within tolerance.
Check air flow and look for fan obstructions.
19
System Overload
Contact Dymax.
100
Emitter Temperature Alarm
>80°C
Reduce obstructions from emitter airflow.
Reduce ambient temperature of emitter.
Reduce Emitter intensity setting.
Check that emitter vent holes are not blocked.
Check vent min clearance distance (1.5 inches).