TROUBLESHOOTING
WARNING:
Only qualified maintenance personnel should attempt the following procedures:
AVERTISSEMENT:
Seulement personnel d’entretien diplomé devrais essayer les procedures suivant.
Problem
Possible Cause
Testing
Corrective Action
Bulb Will Not
Ignite
Improper connections
Visually inspect all input/output
connections (i.e. Power Cord,
Bulb).
Secure all connections.
Bulb beyond useful life of
2,000 hours
Replace with a new Bulb/
Reflector Assembly. Reset the
Bulb Hour Meter and re-test unit.
Replace the Bulb/Reflector
Assembly if required
(typical life = 2,000 hours).
Main line Fuse blown
(nothing in unit operates)
Remove the Fuse from the
Power Receptacle and check it
with an Ohmmeter.
Replace the Fuse if defective.
Low Output
Intensity
or
Fails To Cure
Adhesive In
Allotted Time
Bulb beyond useful life
Use a Radiometer (DYMAX
ACCU-C
AL™ 50 or equivalent)
to measure output intensity.
Replace the Bulb/reflector
assembly if beyond useful life
(typical = 2,000 hours).
Transmission loss in light
guide too great
Compare Lightguide output
against new Lightguide (or use
the DYMAX Lightguide
Simulator) to determine
transmission loss.
Replace the Lightguide.
Contaminants on
Lightguide
Visually examine ends of
Lightguide for contaminants.
Clean with isopropyl alcohol
(or equivalent). Heavy
deposits on liquid Lightguides
may be removed with a razor
blade. Replace Lightguide if it
can not be cleaned.
Bulb/Reflector Assembly
not installed properly
Visually check to make sure the
Bulb/Reflector Assembly is
seated flush in the Bulb Mount
Assembly (any error in
installation could cause a low
output).
Properly install Bulb/Reflector
Assembly.