14
BlueWave
®
50 Light-Curing Spot Lamp
MAINTENANCE
The BlueWave
®
50 was designed to operate with minimum maintenance. Follow the schedule below to assure
top performance from the unit.
LIGHTGUIDE
Clean the ends of the Lightguide monthly or as required. The ends of the Lightguide should be kept clean to
transmit as much light as possible. Cured adhesive can be removed with a razor blade. Avoid sharp bends with
the Lightguide since this reduces light output and damages guide.
FAN FILTERS
The external Fan Filters should be inspected and cleaned periodically to prevent dust buildup from affecting airflow
through the unit. Spare Fan Filters are provided with each unit and with replacement Bulbs. The Fan Filters are
washable and may be reused. Remove the Fan Filter by removing the snap-on cover from the rear of each grill.
Figure 13.
Exhaust Fans with Fan Filters Removed
Figure 14.
Disassembled Fan Filter
FUSE REPLACEMENT
The unit has two Fuses that are installed in the Power Receptacle. To remove the Fuses, unplug the unit and
remove the Fuse Holder with a small screwdriver. Remove the Fuses from the Fuse Holder and install new
Fuses. Replace the Fuse Holder into the Power Receptacle.
Figure 15.
Power Receptacle
Figure 16.
Fuse Holder
Figure 17.
Fuses Removed from Holder