26
RMO40TD DV-Win Software
Test current
It is possible in DV-Win software to enter test current values in two ways. The first way is drop down
menu where you can choose test currents like they are set in device (5 mA, 10 mA, 50 mA, 100 mA, 500
mA, 1 A, 2 A, 5 A, 10 A, 15 A, 25 A, 40 A
). The second option is to choose “Tools”, then “Options” and
then by checking “Test current step 1 A” and OK button it is possible to set all test current from 4 A to 40
A in 1 A step.
Resistance compensation
Resistance compensation is a feature which enables displaying of measured resistance compensated to
reference temperature. This feature is available when controlling RMO40TD from the PC. Two additional
columns of resistance results are displayed in the data table (grid), denoting the compensated resistance
values.
Default temperature values are 25 [
C] for the device under test temperature and 75 [
C] as a reference
temperature. The operator can change resistance compensation parameters such as temperatures or
winding material.