Manual RMO-H
M-RH00NN-300-EN
15
Once a test is finished, the RMO-H automatically changes to the
Result
menu to display the test results.
The
Result
menu provides the result memory location, measured resistance value, measured test
current and voltage, as well as the parameters set for the test (set test current).
3.5.1 Test Results with Rmax Function Enabled
If the
Rmax
menu is set to option
ON
, once the test is finished, RMO-H automatically changes to the
Rmax
to display the test results. In case the measured resistance is equal to or greater than maximum
assigned value, the result will be displayed in “red” color, as presented in the figure below:
In case the measured resistance is lower than set R
max
value the result will be displayed in “green” color,
as presented in the figure below:
The default value for R
max
is 100
µΩ. When the device is turned off and then turned
on, the RMO-H remembers the last saved setting of an R
max
value and a status.
Figure 3-13:
The
Rmax
showing
the test results
Figure 3-12:
The
Result
menu
showing the test results
obtained in Single test
Test current reached
Memory position that holds the results
Measured resistance
of test object
Figure 3-14:
The
Rmax
showing
the test results
Test current set
Measured voltage drop
at test object
101.4
µΩ
101.4
µΩ