Usage
7
Operation
CAUTION
Risk of burns from hot surfaces -
Operate the unit only with the cover
closed
During operation, the interior of the unit
heats up and the surface of the com-
pressor unit becomes hot.
❯
Operate the unit only with the cover
closed.
❯
Do not make any adjustments to the
open unit without wearing protective
gloves.
❯
Allow surfaces to cool down before
performing operating or maintenance
work.
7.1
Switching the unit on/off
9
9
On/Off switch
❯
The unit is switched on by pressing the on/off
switch (9).
The unit starts up and the pressure vessel is fil-
led. When the cut-off pressure is reached the
unit switches off automatically.
❯
The unit is switched off by pressing the on/off
switch (9) again.
Maximum operating
pressure:
8 bar
Switch-on/cut-off pres-
sure:
5 /7 bar
7.2
Start-up cycles
The motors in the compressor stations are
designed for 10 starts/stops per hour. More fre-
quent switching will lead to increased wear.
7.3
Setting the pressure reducer
The compressed air connection to the pressure
reducer has a constant pressure of 1 bar (set at
the factory).
10
8
9
5
5
Pressure reducer
8
Quick-release coupling
9
On/Off switch
10
Pressure gauge
The pressure reducer (5) regulates the desired
working pressure at the quick-release coupling
(8).
The pressure reducer (5) can be adjusted. The
maximum constant operating pressure is 5 bar.
Adjusting the pressure reducer:
The constant operating pressure can be adjusted
by turning the pressure reducer (5). The pressure
can be read off at the pressure gauge (10).
❯
To increase the supply pressure:
Turn the
pressure reducer (5) clockwise towards "+".
❯
To decrease the supply pressure:
Turn the
pressure reducer (5) anti-clockwise towards
"-".
Usage
0715100100L02 2111V005
19
EN