With the aid of a computer, monitor and imaging
software, this digital system can be used to cre-
ate and store images and videos. The following
accessories must always be used: a spacer (not
for the Cam interchangeable head) and hygienic
protective covers.
2.3
Improper use
Any use of this appliance / these appliances
above and beyond that described in the Installa-
tion and Operating Instructions is deemed to be
incorrect usage. The manufacturer cannot be
held liable for any damage resulting from incor-
rect usage. The operator will be held liable and
bears all risks.
Do not operate the device in any rooms in con-
taining flammable mixtures, e.g. in operating the-
atres.
Do not use the camera directly on the eye.
2.4
Indication
Intraoral camera with Cam interchangeable
head
The images aid diagnosis, patient communica-
tion and patient instruction, and are used for
instruction and documentation purposes.
Intraoral camera with Proof interchangeable
head
The intraoral camera with Proof interchangeable
head is intended to assist with the detection and
diagnosis of caries.
Intraoral camera with Proxi interchangeable
head
The intraoral camera with Proxi interchangeable
head is a diagnostic aid for recognition of
approximal caries above the gingiva and for mon-
itoring of the progress of this type of lesions.
2.5
Contraindication
Intraoral camera with Proof interchangeable
head
Large-scale tooth restorations can falsify the dis-
played caries value.
Intraoral camera with Proxi interchangeable
head
The Proxi head is not designed for use on artifi-
cial teeth, teeth with crowns or teeth with exces-
sively large fillings. The device only works
together with natural enamel in the mouth of the
patient. After extraction, teeth can no longer be
analysed with the Proxi head.
2.6
General safety information
❯
Always comply with the specifications of all
guidelines, laws, and other rules and regula-
tions applicable at the site of operation for the
operation of this unit.
❯
Check the function and condition of the unit
prior to every use.
❯
Do not convert or modify the unit.
❯
Comply with the specifications of the Installa-
tion and Operating Instructions.
❯
The Installation and Operating Instructions
must be accessible to all operators of the unit
at all times.
2.7
Combining devices safely
Take care when connecting units together or to
parts of other systems as there is always an ele-
ment of risk (e.g. due to leakage currents).
❯
Only connect units when there can be no
question of danger to operator or to patient.
❯
Only connect units when it is safe to do so and
when there is no risk of damage or harm to the
surroundings.
❯
If it is not 100% clear from the unit data sheet
that such connections can be safely made or if
you are in any doubt, always get a suitably
qualified person (e.g. the manufacturer) to ver-
ify that the setup is safe.
❯
Observe the specifications of IEC 60601
‑
1
(EN 60601
‑
1) when connecting the appliance
with other appliances, e.g. a PC system, both
in and outside the patient environment.
❯
Only connect peripheral units (e.g. computer,
monitor, printer) that conform at least to the
requirements set out in IEC 60950
‑
1
(EN 60950
‑
1).
A copy of the system manufacturer‘s
declaration in accordance with Article 12
of Directive 93/42/EEC can be found in
our download section at www.duerrden-
tal.com (document no. 9000-461-264).
Important information
2109100026L02 1907V001
5
EN
Summary of Contents for VistaCam iX HD Smart
Page 2: ......