11 Cleaning
11.1 Cleaning the optical element
The optical element is located partly in the inter-
changeable head and partly in the handpiece.
NOTICE
Damage of the optical element from
incorrect cleaning
❯
Only use the cleaning set for VistaCam
optical element. Disinfectant residues
soil the optical element.
❯
Clean the window of the optical window of the
interchangeable head from outside using the
microfibre cloth with a drop of VistaCam optical
element cleaner or alcohol.
❯
If particles can still be seen on the image, dry
clean the interchangeable head from the inside
with compressed air or with a foam rod (from
the cleaning set).
❯
Dry clean the lens in the handpiece with com-
pressed air or with a cotton bud (from the
cleaning set).
11.2 Storage box
Clean the surface of the storage box and the
internal shelf in the event of contamination or visi-
ble soiling and disinfect.
Use the following cleaning materials for the stor-
age box:
ü
FD 366 sensitive disinfectant for sensitive sur-
faces
Use the following cleaning materials for the shelf:
ü
FD 350 disinfectant wipes
❯
Clean the surface of the storage box and the
shelf with a dampened, soft, lint-free cloth.
❯
Disinfect the storage box with spray disinfec-
tant on a soft, lint-free cloth. Comply with the
operating instructions for the disinfectant when
doing this.
❯
Disinfect the shelf using a disinfection wipe.