Operation
19 Maintenance
To ensure proper operation and maintain the service life of the appliance, it is necessary to
carry out cleaning and maintenance work diligently�
Inadequately or not carried out maintenance work can cause premature defects that are not
covered by the warranty�
Wear protective equipment to avoid any risk of infection (e�g� liquid-tight protective gloves,
protective goggles, face mask)�
19.1 Maintenance schedule
Service interval
Maintenance work
Daily, before commencing work
Perform a Helix test, following your country-specific regulations�
Every time the device is switched
on
Automatic self-diagnosis
Before each sterilisation process
– Check the chamber, door seal and carrier for contamination,
deposits and damage�
– Carefully clean, disinfect and maintain the parts for
sterilisation
Weekly, before commencing work
– Perform a vacuum test with the device cold and dry�
– Wipe clean and dry the covers of the waste water and fresh
water containers on a regular basis�
Monthly
As required
Clean and disinfect accessory parts (e�g� tray, tray handle)�
If there is soiling
– Carry out complete cleaning (see "19�2 Complete cleaning")�
– Clean or replace the water filter (see "19�5 Clean or replacing
– Clean the tray handle and trays�
– Clean and dry the sterilisation chamber�
– Clean the door seal�
After 1000 cycles or 1 year
Replace the door seal (see "19�3 Cleaning or replacing the door
seal")�
After 2000 cycles or 2 years
Change the sterile air filter (see "19�6 Changing the sterile air
filter")�
After 4000 cycles or 2 years
– Arrange for
maintenance
to be performed by a trained
service technician a few weeks before repeated performance
qualification (revalidation)�
Any modifications to the system
must be subject to evaluation and may require a renewed
performance qualification.
– Renewed
performance qualification
to maintain the efficacy
of the process performed by correspondingly trained expert
personnel or a test laboratory, see also EN ISO 17665
1:2006, DIN SPEC 58929; comply with country-specific
regulations�
Every 2 years
Have electrical safety testing performed, e�g� in accordance
with DIN VDE 0701 0702, and ensure compliance with coun-
try-specific regulations�
Every 4 years
Arrange for replacement of the safety valve� Notify customer
service�