Operation
Avoid the formation of stains.
Proper instrument preparation can prevent resi-
due from being loosened from the load under
steam pressure during sterilisation�
Loosened remains of dirt can clog the filters,
nozzles and valves of the unit and settle as
stains and deposits on the instruments and the
chamber�
A single rusting instrument can be enough to
cause external rust to form on the other instru-
ments or in the unit�
All steam-conducting parts of the appliance are
made of non-rusting materials� This prevents the
formation of rust being caused by the unit�
i
Only use high-quality distilled or demineralised
water�
The extent of stains forming on the instru-
ments can also depend on the quality of the
fresh water�
Replacing the water
Manual emptying of the water containers
1
2
1 Connection for manual fresh water
emptying
2 Connection for manual waste water
emptying
NOTICE
The water in the waste water con-
tainer is contaminated.
i
Take corresponding precautionary
measures when draining it�
i
Carry out disposal in accordance with
national regulations�
i
Place an empty container at the ready with a
capacity of at least 5 litres�
i
Place one end of the drain hose into the con-
tainer provided�
i
Push the drain hose onto a quick release cou-
pling until it noticeably engages to drain the
corresponding water container�
i
Allow all of the fresh water to drain out into
the container, then empty the container�
19.2 Complete cleaning
For problem-free operation of the unit, correct
cleaning and maintenance are necessary�
Only carry out cleaning measures when the
appliance is cold�
i
Switch on the unit and open the door�
i
Switch off the unit�
i
Unplug the mains plug�
i
Use a soft, lint-free cloth�
NOTICE
Leaks in the unit.
Abrasive cleaners can cause leaks�
i
Do not use abrasive cleaners�
i
Do not use hard objects, such as pot scourers
made of metal or steel brushes� Similarly,
avoid using abrasive or scratchy cleaning
sponges�
i
Use cleaning agents that are free of chlorine
and acetic acid�
i
First soak the cloth with cleaning alcohol or
methylated spirits, and then wipe off the dirt
or contamination�
i
In the case of stubborn dirt in the chamber or
on the carrier, use a mild, chlorine-free clean-
ing agent that is suitable for stainless steel
and has no abrasive properties (ph value:
5-8)�
i
Use a neutral liquid cleaner for housing parts
and the door seal�
i
Make sure that no cleaning agents get into the
pipes that come out from the chamber�
Clean the cover
i
Wipe clean and dry the covers of the waste
water and fresh water containers on a regular
basis�