6.2 Alarm Codes
Code Message
Description
Reset
procedure
AL 001
AL 002
AL 003
AL 100
AL 101
AL 102
AL 103
AL 104
AL 105
AL 200
AL 201
AL 202
AL 203
AL 300
AL 301
AL 302
AL 303
AL 304
AL 305
AL 306
AL 307
AL 308
AL 309
AL 310
AL 311
AL 401
AL 402
Reset pressed
Blackout
Water illing time
Chamber probe fault
Generator probe fault
Bandle probe fault
Pressure transducer
Door microswitch
Safety microswitch
Chamber safety cut-off
Generator safety cut-off
Bundle safety cut-off
Safety valve fault
Hardware error
Reboot the system
Valve drive voltage
Mains frequency error
Watchdog trip
Communication error with Main Board
Reboot system
Hardware error
Hardware error
Hardware error
Hardware error
Hardware error
Hardware error
Hardware error
Generator temperature low
Band temperature low
The reset button was pressed
A blackout has occurred
The time to ill the feed tank is too long
The chamber probe reading circuit is faulty
The generator probe reading circuit is faulty
The pipe bundle probe reading circuit is faulty
The pressure tansducer is faulty
The door microswitch has opened unexpectedly
The safety microswitch has opened unexpectedly
The chamber has exceeded the safety temperature
The generator has exceeded the safety temperature
The pipe bundle has exceeded the safety temperature
The pressure has exceeded the maximum trip threshold
of the safety valve
There is no synchronisation for managing AC loads and
checking the mains frequency
There is no valve drive voltage due to a PTC trip
The mains frequency is < 45 Hz or > 65 Hz
Program execution has crashed
There is a communication fault between the main board and
the CPU board
Hardware fault on the main board
Hardware fault on the main board
Hardware fault on the main board
Hardware fault on the main board
Hardware fault on the main board
Hardware fault on the main board
The cycle has incorrectly been closed – Software error
Generator failure
Band failure
A/B
A/B
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
A
6.3 Messages
Code Phase
Message
Msg001
Msg002
Msg003
Msg004
Msg005
Msg008
Msg009
Msg010
Msg011
Msg012
Msg013
Msg014
Msg016
Msg017
Msg018
Feed tank empty
Feed tank full
Incorrect Pin
Poor water quality
Door open
Chamber temperature over limit
Serial number inconsistency between CPU and MB
Lubricate the automated system of the door
Replace the Bio-X ilter
Conduct maintenance
System date incorrect - Check
Do vacuum test
PIN too short
PIN already exists
Firmware version incompatible – contact Technical Support
Standby or cycle start
Standby or cycle start
Pin entry
Water loading
Cycle start
Vacuum Test start
Machine switch-on
Machine switch-on
Machine switch-on
Machine switch-on
Incorrect date
Do vacuum test
PIN too short
Duplicate PIN
MB FW revision incompatible with CPU FW