COD.988250035_REV.0.0_ED.06/2020
56
EN
NOTE
To prevent the formation of limescale stains, rinse with
deionised or distilled water if possible. Always dry the
instruments if you use hard tap water.
For handpieces (turbines, counterangles, etc.), in addition
to the above, use special equipment that will clean them
properly inside (which may also include lubrication).
NOTE
When the sterilisation program has inished, remember to
lubricate the handpiece internal mechanisms with special
oil. These precautions will ensure that the lifetime of the
instrument is not reduced.
CAUTION
Refer to the instrument/material manufacturer’s instructions and
make sure that it is compatible before treating it in the autoclave.
Strictly follow the usage methods for detergents and disinfectants
and the operating instructions for automatic washing and/or
lubrication equipment.
For textiles, (or porous materials in general), such as lab coats,
napkins, caps and similar, wash them thoroughly and dry them
before treating them in the autoclave.
NOTE
Do not use strong detergents that contain chlorine and/or
phosphates. Do not bleach with chlorine-based products.
These components could damage the tray supports, the
trays and any metal instruments in the sterilisation chamber.
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
02
03
04
01
xx µS/cm
Emergency manual filling
- Insert the part of the hose with the quick coupling into the hole
(9 fig. 1)
- Insert the funnel into the other end of the hose and pour in a
maximum of 2 litres of distilled water.
-
In case of connection to demineralization system (18 fig. 2)
the loading tank is filled automatically.
4.6.2 Draining the discharge tank
- Insert the end of the hose without the quick coupling into a
container to collect the drained water.
- Insert the quick coupling into the hole
(7 fig. 1)
and position the
container under the autoclave to drain the tank.
- Drain the water in compliance with the waste disposal
regulations in force.
- Possibility to use the drain tank outlet, placed at the back
(15 fig. 2).
4.6 Filling and draining water
Volume of water used for a sterilisation cycle/load with the highest
consumption: 700ml
Minimum water tank illing level: 1.1 litres
4.6.1 Filling the feed tank
- Only use water that complies with the technical speciications on
page 68.
- The display shows MSG001 (empty load tank), ill the feed tank in
one of the following two ways:
Automatic filling from the front
- Insert the part of the hose with the quick coupling into the water
illing itting
(6 fig. 1)
and put the other end into the water container.
- Select the “water illing” menu
- Press the icon “PLAY WATER FILLING” to start the water
illing pump.
- Filling stops automatically when the water reaches the maximum
level in the feed tank .
- Press the icon “STOP” to voluntarily stop illing.
REAR AUTOMATIC FILLING
In case of connection to the demineralization system
(18 fig. 2)
the
illing of the charging tank is automatic at the beginning and end of
the sterilization cycle. During the water loading the machine cannot
perform sterilization cycles and tests.
01 Select the “Water illing” menu
02 Select “Start” to start illing
03 Select “Stop” to stop illing
04 The cursor position indicates the water quality
N.B. Filling will stop automatically when the maximum level has
been reached, and you will return to the Home screen