D-FL 200 / Rev. 6
29
______________________________________________________________________________________________________
Industrie Elektronik GmbH & Co KG
R23 Amplitude
min.
Minimum signal amplitude, values below this limit are not
evaluated. If this appears 30 seconds in a row, the error
detection is launched (E16-E19),
(0-2000), default 350
R24
amplifier target
Set point for the automatic gain control,
constant 1900
R25
amplification f..
Amplification factor/ manual/ automatic; values
between 1 and 512 will set the variable amplifier to a fixed
amplification, values greater than 512 enable the automatic
gain control (0-1000),
default 600
R26
assessment level
Assessment level, used for searching of the trigger point,
constant 10
input temperature
mA input for temperature
input pressure
mA input for absolute pressure
R27
error detection
Error detection system, enabled=1, disabled=0; single error
evaluations can be disabled by setting the value or the sum
of values: 2
Æ
E12-E15; 4
Æ
E16-E19; 8
Æ
E20-E23; 16
Æ
E24-
E27; 32
Æ
E28-E31; 64
Æ
E32-E33
R28
zero suppress
Zero suppress of gas flow velocity (0-3m/s) (10ft)
R29
sensor offset
Sensor offset , (0-600mm), (corr. to 23,6in),
default 0
R30
noise limit
noise limit, signals with a higher noise component are not
evaluated. If this appears longer than 30 seconds in a row,
the error detection is launched (E24-E27) (0-2000), default
450
R31
ultras. speed NT
Ultrasonic speed at 0°C (100-500m/s),
default 332 m/s (corr. to 1089 ft/s)
R32
calibration time
calibration interval (0-720h), disabled=0,
default 24h
R33
normal/reverse
inverse mode for the velocity measurement (0=normal
direction, 1= inverse), default 0
10.3.3. Third Page / Measuring
On this page the actual reading of the D-FL 200 system can be requested. This is done once if the
"Request“-Button is clicked. If a permanent display is necessary the automatic refresh box has to be
activated. In this case the actual readings of the system are renewed every 5 seconds.