D-FL 200 / Rev. 6
27
______________________________________________________________________________________________________
Industrie Elektronik GmbH & Co KG
(Fig. 23) Second page / Parameter
Parameter
Parameter name
Meaning
Code
1
Store release code for parameterisation with
‘STO-Key’ code = 100
R01
integration time
Integration time (1-180s)
R02
limit value
Limit value 1,in relation to the standard volumetric flow (1-
9999x10
3
Nm
3
/h) (1Nm
3
/h =35,3Nft
3
/h)
R03
measuring path
Measuring path length (0-30m) (1m=3,28ft)
R04 Angle
Angle of installation (0°-80°)
R05 Cross-section
Stack cross section (0-99,99m²) (1m²=10,76ft²)
R06 Corct-factor
Correction factor (0-5) default 1
R07 Offset
Zero point offset (-2m/s…+2m/s) default 0
R08
Span mA .out1
Measuring range end value of the mA output for
the standard volumetric flow measurement
(1-99,99x10
3
Nm³/h) (1Nm
3
/h =35,3Nft
3
/h)
R09
Span mA .out2
Measuring range end value of the mA output for
the gas velocity measurement (0-99,99 m/s)