75
CHART NO. 3 3
SYMPTOM-Refrigerator is TOO COLD
Remember –VACCINE IS DESTROYED IF FROZEN
Check for the following possible faults by first looking at the top of the chart. If this is
found not to be the fault then move to the next fault in the direction of the arrows.
Possible fault
Fault finding & Repair
task No.
Page no
Movable Separator replace or repair seperator
Between refrigerator and
Freezer damaged, incorrectly
Positioned or removed.
(Applies only to some types
Of refrigerator cabinets,
E.g. Electrolux RCW 42)
‘Continuous running’,” draw down” Switch the
Or “freezer” switches on. Refrigerator back
(Applies only to some types on to normal mode.
Of refrigerator with these
additional switches)
Thermostat faulty D.2.1 67
Summary of Contents for VC150
Page 1: ......
Page 3: ...3 SECTION I ...
Page 4: ...4 ...
Page 24: ...24 MODULE AND BATTERY WIRING DIAGRAMS SYSTEM WIRING DIAGRAM ...
Page 25: ...25 SYSTEM WIRING DIAGRAM GENERAL LAYOUT DIAGRAM ...
Page 26: ...26 GENERAL LAYOUT DIAGRAM ELECTRICAL CIRCUIT DIAGRAM ...
Page 27: ...27 ELECTRICAL CIRCUIT DIAGRAM REFRIGERATION CIRCUIT DIAGRAM ...
Page 28: ...28 REFRIGERATION CIRCUIT DIAGRAM ...
Page 29: ...29 SECTION II ...
Page 38: ...38 ...
Page 39: ...39 ...
Page 42: ...42 ...
Page 44: ...44 ...
Page 50: ...50 ...
Page 124: ...124 ANNEX 1 COMPONENT LAYOUTS OF SOME COMMONLY USED SOLAR REFRIGERATOR ...
Page 125: ...125 ...
Page 126: ...126 ...
Page 127: ...127 ...
Page 128: ...128 ...
Page 129: ...129 ...
Page 130: ...130 ANNEX II ADDITIONAL EPI TOOL KIT NEEDED FOR PHOTOVOLTAIC REFRIGERATOR REPAIR ...
Page 136: ...136 ...
Page 137: ...137 ...