30
9000-608-21/30
2013/10/21
Problem
Possible cause
Solution
15. X-ray exposure
has stripes on
image
• Image plate in the cassette was
exposed by natural or scattered
X-ray radiation
• Only use the image plate within
24 hours of data erasure
• Image plate in the cassette soiled
or scratched
• Inform your service technician
16. Warning mes-
sage: "Too much
ambient light"
• CR 43 exposed to too much light
• Darken the room
• Turn the CR 43 so that the light
does not fall directly onto the
input unit
17. Pale stripes in
scanning window
• Too much stray light entered dur-
ing scanning
• Darken the room
• Turn the CR 43 so that the light
does not fall directly onto the
input unit
18. Narrow images
with light and
dark streaks ap-
pear in addition
to the correct
image
• Ambient light entered during
scanning
• Darken the room
• Turn the CR 43 so that the light
does not fall directly onto the
input unit
19. Blue LED out
• No connection between CR 43
and PC
• Activate the CR 43 via the soft-
ware
20. Red LED lit
• Device fault:
Urgent need for action by the
operator
• Switch off the CR 43 for approx
10 seconds
• Restart the CR 43
• If the CR 43 does not start, inform
the service technician
EN