If
cursor velocity is
= 50 mm / sec.
50 mm / sec
Step =----------------------------- = 0,05 mm / samples
1000 samples/ sec
The overmentioned resolutions depend also on the sampling rate and A/D converter
capability of the electronics that is getting the transducer signal.
10. ELECTRICAL
CONNECTIONS
In this chapter you are given information about electrical connections of series PC
transducers.
It has been listed into: power supply, analog output and serial output.
Before feeding transducer check that the connections have been done according to
following instructions:
In the case of interconnections longer than 5 meters, use conductors with diameters
greater than 1 mm. In any case. It is to use only 100% shielded cables.
In the following text, the indications refer to the cable leads (option C, output with cable); in
brackets (…) we have indications about connector pins (option P, connector output).
Example: WHITE (PIN3) means: WHITE output with cable, while (PIN3) refers to output
with connector.
Relationship between cursor translation velocity and FS, for significant measuring step
0,01
0,1
1
10
Step
Velo
city
m/s
0,05 mm
0,0125
0,025
0,05
0,1 mm
0,025
0,05
0,1
1 mm
0,25
0,5
1
2 mm
0,5
1
2
5 mm
1,25
2,5
5
10 mm
2,5
5
10
FS from 1500 to 3000 mm
FS from 350 to 1500 mm
Fs up to 350 mm