Drucker Diagnostics 614VET Operator'S Manual Download Page 9

Page | 9 Drucker Diagn1-814-692-7761 - [email protected]

 

O

PERATION

 

NOTE: Follow the initial setup procedure on page 6 before initial operation.  

1.

 

Plug the centrifuge into an approved 115 Volt A.C., 60 Hz. outlet. 

2.

 

Push the ‘OPEN/STOP’ button and then open the lid. 

3.

 

Place the test tube samples into the tube holders. Be sure to 

follow the rules for balanced loads. 

4.

 

Close the lid and turn the lid knob clockwise to its complete stop 

position. The ’LATCHED’ indicator light should turn on to indicate 

that the latch is closed properly. If the lid knob is not completely 

latched, the ‘LATCHED’ indicator light will not turn on and the 

centrifuge will not operate! 

5.

 

The timer has been set to a preset time of two (2) minutes. To 

display or change this time setting, refer to page 8. 

6.

 

Turn on the machine by pushing the ‘START’ button on the 

control panel. 

7.

 

The centrifuge should begin to spin. The ‘RUNNING’ indicator 

light should illuminate. 

IF A PROBLEM IS FOUND DURING A SPIN THAT REQUIRES THE 

CENTRIFUGE TO SHUT DOWN, PRESS THE ‘OPEN / STOP’ BUTTON! 

8.

 

The ‘RUNNING’ indicator light will begin to flash when one 

minute remains. 

9.

 

After time has elapsed, the ‘RUNNING’ indicator light will 

extinguish and the rotor will slow to a complete stop. 

10.

 

The ‘UNLOCKED’ indicator light will illuminate, and the locking 

mechanism will disengage allowing entry into the rotor chamber. 

If it does not, refer to page 13 on troubleshooting. 

11.

 

Turn the lid knob counterclockwise and open the lid. 

12.

 

Remove the samples. 

13.

 

If the machine re–locks before the samples are removed, press 

the ‘OPEN/STOP’ button to unlock the lid for an additional fifteen 

(15) seconds. 

 

 

Summary of Contents for 614VET

Page 1: ...614VET Operator s Manual P N 03 0 0002 0229 Rev C...

Page 2: ...5 Initial Setup Procedure 6 Control Panel 7 Verifying The Preset Time 8 Changing The Preset Time 8 Operation 9 Spinning Balanced Loads 10 Rotor Removal And Installation 10 To remove the rotor 10 To i...

Page 3: ...e holders inside machine MODEL DESCRIPTION The Model 614VET is a continuous duty electronically controlled Fixed Angle centrifuge with a lid safety interlock system The unit is controlled by an electr...

Page 4: ...flow design that prevents overheating of samples Heavy gauge steel construction for safety and durability Lid safety switch that prevents the centrifuge from operating unless the lid is closed and la...

Page 5: ...uipment is present 2 Choose a setup location which meets the following criteria a A clearance height of 20 is required to open the lid b The clearance envelope is the space around the centrifuge which...

Page 6: ...the rotor and verify that they are seated properly 6 Close the lid Rotate the lid knob clockwise to its complete top position The LATCHED indicator light should be illuminated If it is not make sure...

Page 7: ...for fifteen 15 additional seconds After the centrifuge has passed this procedure it is ready for operation CONTROL PANEL RUNNING Lights up when the machine is in operation power is being applied to t...

Page 8: ...CHANGING THE PRESET TIME NOTE Your centrifuge must be plugged in a Push the OPEN STOP button to disengage the lock and then open the lid b Push and hold the START and OPEN buttons for approximately th...

Page 9: ...of two 2 minutes To display or change this time setting refer to page 8 6 Turn on the machine by pushing the START button on the control panel 7 The centrifuge should begin to spin The RUNNING indicat...

Page 10: ...plug the centrifuge from the electrical outlet to eliminate the possibility of electrical shock or other injury 2 Remove the test tube holders 3 Remove the nut in the center of the rotor by turning it...

Page 11: ...e safety guidelines of your laboratory to properly clean up and or dispose of materials in the event that a substance known to be potentially toxic radioactive or contaminated with a pathogenic microo...

Page 12: ...in water or other cleaning solutions as this will cause damage and void the warranty 3 ONLY isopropyl alcohol soap and water or a 10 5500 PPM bleach solution should be used for cleaning and disinfect...

Page 13: ...replace it if necessary The centrifuge does not run o Check the electrical outlet o Make sure the lid latch is turned completely clockwise to its stop position When the lid is closed properly the lat...

Page 14: ...t to the desired length Check the run preset by following the instructions on page 8 If the preset is not the desired length follow the procedure on the same page to change the run preset time EMERGEN...

Page 15: ...14VET has a true 0 RPM lid locking system The lid safety locking system keeps the lid locked at all times even during power failure and requires that the rotor be at rest in order to unlock the lid Th...

Page 16: ...d 7751043 Circuit Breaker 7760002 Power cord 03 1 0008 0016 Pawl latch lid 03 1 0008 0009 Knob latch lid 7724071 Hinge friction 7732018 Seal lid gasket 7713032 125mm Tube Holder Black 02 002 1 0027 Li...

Reviews: