ROUBLESHOOTING

 

Note: The latch must be turned completely clockwise to its stop position 

in order for the centrifuge to operate. 

The rotor does not spin 

freely. 

o

 

Make sure nothing has fallen into the 

rotor chamber. 

o

 

If there is nothing obstructing the rotor, 

the rotor may be damaged. contact 

Drucker Diagnostics for further 

assistance. 

Excessive noise when 

the machine is 

running

o

 

Check to see that the load is balanced. 

o

 

Make sure that nothing has fallen into 

the rotor chamber. 

o

 

Make sure that the nut in the center of 

the rotor is tight. 

o

 

Have a technician test the motor and 

replace it if necessary. 

The centrifuge does 

not run. 

o

 

Check the electrical outlet. 

o

 

Make sure the lid latch is turned 

completely clockwise to its stop 

position. When the lid is closed 

properly, the latch light on the control 

panel will illuminate. 

o

 

Check the circuit breaker switch at the 

bottom left of the machine. If the switch 

is white, the breaker has tripped. 

Contact Drucker Diagnostics for further 

assistance. 

o

 

The printed circuit board may be 

damaged. Have a technician test and 

replace the circuit board if necessary. 

The latch light does 

not come on when the 

lid is closed. 

o

 

Make sure that the unit has power. 

o

 

Make sure the lid latch is turned 

completely clockwise to its stop 

position. The latch makes contact with a 

switch underneath the front top of the 

cabinet. If this switch is not activated, 

the light will not turn on and the 

machine will not run. 

 

Summary of Contents for 614VET

Page 1: ...614VET Operator s Manual P N 03 0 0002 0229 Rev C...

Page 2: ...5 Initial Setup Procedure 6 Control Panel 7 Verifying The Preset Time 8 Changing The Preset Time 8 Operation 9 Spinning Balanced Loads 10 Rotor Removal And Installation 10 To remove the rotor 10 To i...

Page 3: ...e holders inside machine MODEL DESCRIPTION The Model 614VET is a continuous duty electronically controlled Fixed Angle centrifuge with a lid safety interlock system The unit is controlled by an electr...

Page 4: ...flow design that prevents overheating of samples Heavy gauge steel construction for safety and durability Lid safety switch that prevents the centrifuge from operating unless the lid is closed and la...

Page 5: ...uipment is present 2 Choose a setup location which meets the following criteria a A clearance height of 20 is required to open the lid b The clearance envelope is the space around the centrifuge which...

Page 6: ...the rotor and verify that they are seated properly 6 Close the lid Rotate the lid knob clockwise to its complete top position The LATCHED indicator light should be illuminated If it is not make sure...

Page 7: ...for fifteen 15 additional seconds After the centrifuge has passed this procedure it is ready for operation CONTROL PANEL RUNNING Lights up when the machine is in operation power is being applied to t...

Page 8: ...CHANGING THE PRESET TIME NOTE Your centrifuge must be plugged in a Push the OPEN STOP button to disengage the lock and then open the lid b Push and hold the START and OPEN buttons for approximately th...

Page 9: ...of two 2 minutes To display or change this time setting refer to page 8 6 Turn on the machine by pushing the START button on the control panel 7 The centrifuge should begin to spin The RUNNING indicat...

Page 10: ...plug the centrifuge from the electrical outlet to eliminate the possibility of electrical shock or other injury 2 Remove the test tube holders 3 Remove the nut in the center of the rotor by turning it...

Page 11: ...e safety guidelines of your laboratory to properly clean up and or dispose of materials in the event that a substance known to be potentially toxic radioactive or contaminated with a pathogenic microo...

Page 12: ...in water or other cleaning solutions as this will cause damage and void the warranty 3 ONLY isopropyl alcohol soap and water or a 10 5500 PPM bleach solution should be used for cleaning and disinfect...

Page 13: ...replace it if necessary The centrifuge does not run o Check the electrical outlet o Make sure the lid latch is turned completely clockwise to its stop position When the lid is closed properly the lat...

Page 14: ...t to the desired length Check the run preset by following the instructions on page 8 If the preset is not the desired length follow the procedure on the same page to change the run preset time EMERGEN...

Page 15: ...14VET has a true 0 RPM lid locking system The lid safety locking system keeps the lid locked at all times even during power failure and requires that the rotor be at rest in order to unlock the lid Th...

Page 16: ...d 7751043 Circuit Breaker 7760002 Power cord 03 1 0008 0016 Pawl latch lid 03 1 0008 0009 Knob latch lid 7724071 Hinge friction 7732018 Seal lid gasket 7713032 125mm Tube Holder Black 02 002 1 0027 Li...

Reviews: