Page | 13 Drucker Diagn1-814-692-7761 - [email protected]
T
ROUBLESHOOTING
Note: The latch must be turned completely clockwise to its stop position
in order for the centrifuge to operate.
The rotor does not spin
freely.
o
Make sure nothing has fallen into the
rotor chamber.
o
If there is nothing obstructing the rotor,
the rotor may be damaged. contact
Drucker Diagnostics for further
assistance.
Excessive noise when
the machine is
running
.
o
Check to see that the load is balanced.
o
Make sure that nothing has fallen into
the rotor chamber.
o
Make sure that the nut in the center of
the rotor is tight.
o
Have a technician test the motor and
replace it if necessary.
The centrifuge does
not run.
o
Check the electrical outlet.
o
Make sure the lid latch is turned
completely clockwise to its stop
position. When the lid is closed
properly, the latch light on the control
panel will illuminate.
o
Check the circuit breaker switch at the
bottom left of the machine. If the switch
is white, the breaker has tripped.
Contact Drucker Diagnostics for further
assistance.
o
The printed circuit board may be
damaged. Have a technician test and
replace the circuit board if necessary.
The latch light does
not come on when the
lid is closed.
o
Make sure that the unit has power.
o
Make sure the lid latch is turned
completely clockwise to its stop
position. The latch makes contact with a
switch underneath the front top of the
cabinet. If this switch is not activated,
the light will not turn on and the
machine will not run.